Operating Instructions — Batches
A batch is a collection of information about the samples to be analyzed. Batches tell the software
the order in which to analyze the samples. For information about importing batches, refer to the
Advanced User Guide
.
Set Queue Options
The queue goes one by one through the list, acquiring each sample with the selected acquisition
method. After all the samples have been acquired, the queue stops and the mass spectrometer
goes into the standby mode. In the standby mode, the LC pumps and some instrument voltages
are turned off.
The user can change the length of time the queue runs after the last acquisition has finished,
before the Analyst software puts the mass spectrometer into the standby mode. For more
information about the other fields in the
Queue Options
dialog, refer to the Help.
1. In the navigation bar, click
Configure
.
2. Click
Tools > Settings > Queue Options
.
3. In the
Max Num Waiting Samples
field, set the maximum number of samples to a
value that is greater than the number of samples that will be submitted to the queue.
4. In the
Max Idle Time
field, type the length of time the queue will wait after acquisition
is completed before going into Standby mode. The preset value is 60 minutes.
If gas cylinders are used, adjust this time to make sure that the gas in the cylinders is
not depleted.
Figure 9-1 Queue Options Dialog
Содержание QTRAP 4500 LC/MS/MS system
Страница 4: ...Revision Log 4500 Series of Instruments System User Guide 4 of 138 RUO IDV 05 0264 A ...
Страница 22: ...Hazards Symbols 4500 Series of Instruments System User Guide 22 of 138 RUO IDV 05 0264 A ...
Страница 36: ...Operating Instructions Hardware 4500 Series of Instruments System User Guide 36 of 138 RUO IDV 05 0264 A ...
Страница 76: ...Operating Instructions Batches 4500 Series of Instruments System User Guide 76 of 138 RUO IDV 05 0264 A ...
Страница 128: ...Service and Maintenance Information 4500 Series of Instruments System User Guide 128 of 138 RUO IDV 05 0264 A ...
Страница 130: ...Troubleshooting 4500 Series of Instruments System User Guide 130 of 138 RUO IDV 05 0264 A ...
Страница 136: ...4500 Series of Instruments System User Guide 136 of 138 RUO IDV 05 0264 A ...
Страница 138: ...Calibration Ions and Solutions 4500 Series of Instruments System User Guide 138 of 138 RUO IDV 05 0264 A ...