
Operating Instructions — Hardware
4500 Series of Instruments
System User Guide
32 of 138
RUO-IDV-05-0264-A
Plumb the Diverter Valve
Plumb the diverter valve for both injector and diverter modes.
Plumb the Diverter Valve in Injector Mode
The diverter valve is a two position, 6 port valve. If you put the valve in Position A (
the sample flows through the external loop. When you switch the valve to Position B
), the sample is injected.
Figure 4-5 Diverter Valve—Injector Mode Position A
Item Description
1
Sample in
2
Waste out
3
Sample loop
4
Mobile phase in
5
To column
1
4
6
5
2
3
1
2
3
5
4
Содержание QTRAP 4500 LC/MS/MS system
Страница 4: ...Revision Log 4500 Series of Instruments System User Guide 4 of 138 RUO IDV 05 0264 A ...
Страница 22: ...Hazards Symbols 4500 Series of Instruments System User Guide 22 of 138 RUO IDV 05 0264 A ...
Страница 36: ...Operating Instructions Hardware 4500 Series of Instruments System User Guide 36 of 138 RUO IDV 05 0264 A ...
Страница 76: ...Operating Instructions Batches 4500 Series of Instruments System User Guide 76 of 138 RUO IDV 05 0264 A ...
Страница 128: ...Service and Maintenance Information 4500 Series of Instruments System User Guide 128 of 138 RUO IDV 05 0264 A ...
Страница 130: ...Troubleshooting 4500 Series of Instruments System User Guide 130 of 138 RUO IDV 05 0264 A ...
Страница 136: ...4500 Series of Instruments System User Guide 136 of 138 RUO IDV 05 0264 A ...
Страница 138: ...Calibration Ions and Solutions 4500 Series of Instruments System User Guide 138 of 138 RUO IDV 05 0264 A ...