Service and Maintenance Information
System User Guide
4500 Series of Instruments
RUO-IDV-05-0264-A
123 of 138
•
Only dampen the wipe or swab slightly when applying water or cleaning solution.
Water, more so than organic solvents, may cause the wipe to deteriorate, leaving
residue on the mass spectrometer.
Prepare for Cleaning
In routine cleaning, clean the curtain plate and the front of the orifice plate. Routine cleaning can
be performed while the mass spectrometer remains under vacuum.
1. Deactivate the hardware profile.
2. Wait at least 30 minutes for the ion source, curtain plate, and orifice plate to cool.
3. Remove the ion source. Be sure to place the ion source in a safe location.
Caution: Potential System Damage. Be careful not to not drop anything into
the source drain when the ion source is removed.
Note:
Mass spectrometers with a NanoSpray
®
ion source may require a full cleaning
for best results. Contact an FSE.
WARNING! Hot Surface Hazard: Surfaces of the ion source become
hot during operation. Let the ion source and vacuum interface
components cool for at least 30 minutes before starting any cleaning
procedures.
Figure 13-2 Source Drain on the Vacuum Interface
Item Description
1
Source drain
1
Содержание QTRAP 4500 LC/MS/MS system
Страница 4: ...Revision Log 4500 Series of Instruments System User Guide 4 of 138 RUO IDV 05 0264 A ...
Страница 22: ...Hazards Symbols 4500 Series of Instruments System User Guide 22 of 138 RUO IDV 05 0264 A ...
Страница 36: ...Operating Instructions Hardware 4500 Series of Instruments System User Guide 36 of 138 RUO IDV 05 0264 A ...
Страница 76: ...Operating Instructions Batches 4500 Series of Instruments System User Guide 76 of 138 RUO IDV 05 0264 A ...
Страница 128: ...Service and Maintenance Information 4500 Series of Instruments System User Guide 128 of 138 RUO IDV 05 0264 A ...
Страница 130: ...Troubleshooting 4500 Series of Instruments System User Guide 130 of 138 RUO IDV 05 0264 A ...
Страница 136: ...4500 Series of Instruments System User Guide 136 of 138 RUO IDV 05 0264 A ...
Страница 138: ...Calibration Ions and Solutions 4500 Series of Instruments System User Guide 138 of 138 RUO IDV 05 0264 A ...