Service and Maintenance Information
System User Guide
4500 Series of Instruments
RUO-IDV-05-0264-A
121 of 138
Clean the Front-End
Clean the mass spectrometer front-end using the routine cleaning method, to:
•
Minimize unscheduled system downtime.
•
Maintain optimum sensitivity.
•
Avoid more extensive cleaning that requires a service visit.
Symptoms of contamination:
Significant loss in sensitivity and increased background noise.
When contamination occurs, perform an initial routine cleaning. Clean up to and including the
front of the orifice plate. If routine cleaning does not resolve issues with sensitivity, a full cleaning
may be necessary.
This section provides instructions for performing routine cleaning without breaking vacuum and
full cleaning under atmospheric pressure, after venting the mass spectrometer.
Note:
Follow all applicable local regulations. For health and safety guidelines, refer to
Chemical Precautions on page 11
Note:
U.S. customers can call 877-740-2129 for ordering information and inquiries.
International customers can visit
www.absciex.com
.
Required Materials
• Powder free gloves (nitrile recommended)
• Safety glasses
• Laboratory coat
• Fresh, high quality water (at least 18 Mohm de-ionized water (DI) or ultra-pure HPLC-grade
water). Old water can contain contaminants which can further contaminate the mass
spectrometer.
• HPLC- or MS-grade methanol, isopropanol (2-propanol), or acetonitrile
• Cleaning solution. Use one of:
• 100% methanol
• 100% isopropanol
• 50:50 acetonitrile:water solution (freshly prepared)
• 50:50 acetonitrile:water with 0.1% acetic acid solution (freshly prepared)
Table 13-2 Tools and Supplies Available from AB SCIEX
Description
PN
Small polyester swab (thermally bonded). Available in the Cleaning kit.
1017396
Small lint-free wipe (11 cm x 21 cm). Available in the Cleaning kit.
018027
Содержание QTRAP 4500 LC/MS/MS system
Страница 4: ...Revision Log 4500 Series of Instruments System User Guide 4 of 138 RUO IDV 05 0264 A ...
Страница 22: ...Hazards Symbols 4500 Series of Instruments System User Guide 22 of 138 RUO IDV 05 0264 A ...
Страница 36: ...Operating Instructions Hardware 4500 Series of Instruments System User Guide 36 of 138 RUO IDV 05 0264 A ...
Страница 76: ...Operating Instructions Batches 4500 Series of Instruments System User Guide 76 of 138 RUO IDV 05 0264 A ...
Страница 128: ...Service and Maintenance Information 4500 Series of Instruments System User Guide 128 of 138 RUO IDV 05 0264 A ...
Страница 130: ...Troubleshooting 4500 Series of Instruments System User Guide 130 of 138 RUO IDV 05 0264 A ...
Страница 136: ...4500 Series of Instruments System User Guide 136 of 138 RUO IDV 05 0264 A ...
Страница 138: ...Calibration Ions and Solutions 4500 Series of Instruments System User Guide 138 of 138 RUO IDV 05 0264 A ...