Digital Blood Pressure Monitor 

 
 
 
 
 
 
 

Model UA-782

   

 

 

 

 

Instruction Manual 

 

 
 
 
 

 

 

 
1WMPD4001900 

Содержание UA-782

Страница 1: ...Digital Blood Pressure Monitor Model UA 782 Instruction Manual 1WMPD4001900 ...

Страница 2: ......

Страница 3: ...7 Attaching The Arm Cuff 7 How To Take Proper Measurements 8 Measurement 8 After Measurement 8 Measurements 9 Normal Measurement 9 Measurement With The Desired Systolic Pressure 10 Notes For Proper Measurement 10 About Memory 11 Recalling Data 11 Clearing Data 11 What Is An Irregular Heartbeat 12 Pressure Bar Indicator 12 WHO Classification Indicator 12 About Blood Pressure 13 What Is Blood Pressu...

Страница 4: ...se The device is for use indoors Precautions Precision components are used in the construction of this device Extremes in temperature humidity direct sunlight shock or dust should be avoided Clean the device with a soft dry cloth Never use thinner alcohol benzine or wet cloth Avoid tightly folding the cuff or storing the hose tightly twisted for long periods as such treatment may shorten the life ...

Страница 5: ...or Irregular heartbeat indicator Diastolic Pressure Pulse Rate Systolic Pressure Pressure Bar Indicator and WHO Classification Indicator 1 5V Batteries R6P LR6 or AA Battery Compartment Battery Cover Display START button Air Socket Air Connector Plug Arm Cuff Air Hose DC Jack MEMORY button ...

Страница 6: ...ement Previous measurements stored in MEMORY Full Battery The battery power indicator during measurement Low Battery The battery is low when it blinks Replace all batteries with new ones when the indicator blinks Unstable blood pressure due to movement during the measurement Try the measurement again Remain very still during the measurement The systolic and diastolic values are within 10 mmHg of e...

Страница 7: ...e 11 When nothing is displayed press the MEMORY button to appear the data number The data number of the most recent measurement and the data are displayed Stored data are automatically displayed in order from the last data 3 Clearing Data Refer to page 11 Remove batteries to clear all data Re install them again 4 Measurement With The Desired Systolic Pressure Refer to page 10 for the measurement w...

Страница 8: ...d batteries only The batteries provided with the device are for testing monitor performance and may have a limited life Remove the batteries if the device is not to be used for a long time The batteries may leak and cause a malfunction The stored data are cleared when the batteries are removed When the AC adapter is used for measurement memories are cleared after pulling out the AC adapter Install...

Страница 9: ...ed on each cuff If your arm size is over 32 cm the Large cuff is proper Arm Size Recommended Cuff Size Order No 22 cm to 43 cm Large cuff UA CUFUBKLA EC 17 cm to 32 cm Medium cuff UA CUFUBKAU EC Small cuff Arm size The circumference at the biceps Note Model UA 782 is NOT designed for using a small cuff Attaching The Arm Cuff 1 Wrap the cuff around the upper arm about 2 3 cm above the elbow as show...

Страница 10: ... Do not measure right after physical exercise or a bath Rest before taking the measurement Try to measure your blood pressure at the same time every day Measurement During measurement it is normal for the cuff to feel very tight Do not be alarmed After Measurement After measurement press the START button to turn off the power Remove the cuff and record your data Note The device has an automatic po...

Страница 11: ...ish to stop inflation at any time press the START button again 3 When inflation is complete deflation starts automatically and the heart mark blinks indicating that the measurement is in progress Once the pulse is detected the mark blinks with each pulse beat Note If an appropriate pressure is not obtained the device starts to inflate again automatically 4 When the measurement is complete the syst...

Страница 12: ...reflect this stress as a higher or lower than normal blood pressure reading and the pulse reading will usually be faster than normal An individual s blood pressure varies constantly depending on what you are doing and what you have eaten What you drink can have a very strong and rapid effect on your blood pressure This device bases its measurements on the heartbeat If you have a very weak or irreg...

Страница 13: ...n The data number of the most recent measurement and the data are displayed 2 If you want to see the previous data press the MEMORY button until getting the data you want 3 The display will shut off automatically one minute after you press the MEMORY button Note If press the START button when recalling data the display shuts off Clearing Data To clear all data stored in memory Press and hold the M...

Страница 14: ... indicator frequently Pressure Bar Indicator The indicator monitors the progress of pressure during measurement WHO Classification Indicator Each six segments of the bar indicator correspond to the WHO blood pressure classification described on the next page The indicator displays a segment based on the current data corresponding to the WHO classification WHO Classification Indicator Severe hypert...

Страница 15: ...gular physical checkups Maintain proper weight Why Measure Blood Pressure At Home Blood pressure measured at a clinic or doctor s office may cause apprehension and can produce an elevated reading 25 to 30 mmHg higher than that measured at home Home measurement reduces the effects of outside influences on blood pressure readings supplements the doctor s readings and provides a more accurate complet...

Страница 16: ...y the mark does not appear Replace all batteries with new ones The cuff is not fastened properly Fasten the cuff correctly You moved your arm or body during the measurement Make sure you remain very still and quiet during the measurement The cuff position is not correct Sit comfortably and still Raise your hand so that the cuff is at the same level as your heart The device does not measure Reading...

Страница 17: ...echnical Data Type UA 782 Measurement method Oscillometric measurement Measurement range Pressure 20 280 mmHg Pulse 40 180 beats minute Measurement accuracy Pressure 3 mmHg or 2 whichever is greater Pulse 5 Power supply 4 x 1 5V batteries R6P LR6 or AA or AC adapter TB 233 Not included Classification Type BF Clinical test According to ANSI AAMI SP 10 1987 EMC IEC 60601 1 2 2001 Memory Last 90 meas...

Страница 18: ...MEMO ...

Страница 19: ......

Страница 20: ...horized Representative Established in the European Community Unit 24 26 Blacklands Way Abingdon Business Park Abingdon Oxfordshire OX14 1DY United Kingdom Telephone 44 1235 550420 Fax 44 1235 550485 A D MERCURY PTY LTD 32 Dew Street Thebarton South Australia 5031 AUSTRALIA Telephone 61 8 8301 8100 Fax 61 8 8352 7409 ...