67
8.6.3.
Attaching the Cuff, Carry holder and Recorder
Do not attach the cuff if the patient has dermatitis, external
wounds, etc.
Remove the cuff and cease use if dermatitis or other
symptoms appear on the patient.
Prevent the air hose from coiling around the neck and body.
Take care when using around infants, as there is a danger
of suffocation.
Insert the connector of the air hose firmly until the end of its
rotation. If the connection is incorrect, it may cause air
leakage and a measurement error.
Note
Attach the cuff in the correct position and wrap around the arm
to measureblood pressure correctly.
Prevent the cuff and air hose from vibrating during
measurement. The recorder measures a delicate change
in the air pressure inside the cuff.
The accessory cuff is an adult cuff for the left arm. If the cuff
size does not fit, purchase an optional cuff.
Arm circumference
Small cuff
15 to 22 cm
0
5.9" to
0
8.7"
Adult cuff
20 to 31 cm
0
7.8" to 12.2"
Large cuff
28 to 38 cm
11.0" to 15.0"
Extra-large cuff
36 to 50 cm
14.2" to 19.7"
Keep the cuff clean.
We recommend that the patient uses the carry holder and belt.
The cuff is not made from natural rubber latex.
Caution