8.5.
Deleting Measurement Data
Purpose of operation and explanation of function
Measurement data is deleted but the settings are not deleted.
The initial settings can be configured using the following methods.
Using switches on the recorder.
Using a
dedicated peripheral
that is connected to the recorder
using the USB cable.
If measurement data is deleted, it cannot be used again.
Backup the data before deletion.
Delete the measurement data for the last patient before
the next patient uses the recorder.
Deleting the data may take around ten seconds.
Do not operate the device whlie the data is being deleted
to ensure that it is deleted correctly.
Procedure of operation using switches
Step 1. Set the AUTO switch to "
OFF
".
Step 2. If the indication is hidden, press the START/STOP or
EVENT switch to return to the waiting mode display.
Step 3. While pressing and holding the START/STOP switch,
press and hold the EVENT switch until DataClear
(after Sleep and Display ) is displayed on the OLED.
Step 4. Select an operation.
If you wish to delete data, press and hold the START/STOP
switch.
Erasing blinks under DataClear on the OLED and the
deletion of data starts. Proceed to step 5 after deletion.
If you keep (do not delete) data, press the EVENT switch
and proceed to step 5.
Step 5. The recorder returns to the waiting mode.
Содержание TM-2441
Страница 20: ......
Страница 36: ...16 4 2 Dimensions 66 24 5 95 Unit mm...
Страница 89: ...69 Air hose Air hose Cuff Belt Carry holder Air plug Air socket Insert Carry holder Belt Tape Clip...