![A&D TM-2441 Скачать руководство пользователя страница 78](http://html1.mh-extra.com/html/aandd/tm-2441/tm-2441_instruction-manual_3078857078.webp)
58
8.4.
S-BPM Programs
The initial settings can be configured using the following methods.
Using switches on the recorder.
Using a
dedicated peripheral
that is connected to the recorder
using the USB cable.
Procedure for operation using switches
Step 1. Set the AUTO switch to "
OFF
".
Step 2. While pressing and holding the / switch,
press and hold the START/STOP switch until
is
displayed on the LCD.
Step 3. Select S-BPM program using the following switches.
Program: OBP , AOBP, HBP, ANBP, ASBP
/ switch ...................... Change the current parameter.
START/STOP switch ........ Decision, next item.
Step 4. Specify each item (
Measurement count
,
Interval time
,
start time
and
start time of alarm
) using the following switches.
/ switch ...................... Change the current parameter.
START/STOP switch ........ Decision, next item, end of settings.
Step 5. After completing settings, S-BPM clock is displayed.
Do not remove batteries while charging the settings.
If batteries are removed, input settings again.
Caution