5.Trigger Setup - 5.2.Memory Trigger
38
1WMPD4004445
Step 3. Up to 18 trigger sources T1 to T18 can be set.
Tap the number of the trigger source to set to display the details screen.
(1)
(2)
(3)
(4)
(5)
(6)
(1) Trigger menu selection:
Switches between the Memory trigger, Start trigger, or Memory block menu
(2) Channel: Selects the TRIGn source channel.
(3) Detection: Selects UP, DOWN, INTO WIN, or OUT WIN for the polarity of the trigger signal.
UP
The trigger is detected when the value exceeds the trigger level (threshold).
DOWN
The trigger is detected when the value is below the trigger level (threshold).
INTO WIN The trigger is detected when the value enters the range of the upper limit
value or lower limit value of the trigger level.
OUT WIN The trigger is detected when the value leaves the range of the upper limit
value or lower limit value of the trigger level.
(4) Threshold: Sets the trigger level (threshold).
For INTO WIN / OUT WIN, there are two settings: the upper threshold and lower threshold.
(5) Filter:
Sets the filter time for noise removal.
(6) Close:
Ends the setting operation and returns to the trigger list.
Содержание RA3100 Omniace
Страница 1: ...RA3100 Omniace Simple Instruction Manual 1WMPD4004445 ...
Страница 73: ...8 Specifications 8 3 Exterior 1WMPD4004445 73 8 3 Exterior 8 3 1 Main Unit Exterior 394 199 410 334 ...
Страница 74: ...8 Specifications 8 3 Exterior 74 1WMPD4004445 MEMO ...
Страница 75: ...Omniace RA3100 Simple Instruction Manual 1WMPD4004445 1st Edition ...