Command mode
The command mode is the function which can perform "output data", "controlling the
scale" and "setting parameters" by a command transmitted from a computer
Caution
Allow at least 500 milliseconds between commands.
16.3.1.
Command List
The following explanation uses "format 1 (
f15 0
)"
Data output
The current weighing data is output.
Template Q
Command
Reply
Selection of mode and unit
Selects the mode and unit. This is the same as the
MODE
key.
Template U
Command
Reply
The scale changes mode and unit.
Zero
The current mass value is set to the zero point.
This is the same as the
ZERO
key.
Template Z
Command
Reply
The mass value becomes zero and the zero point mark is displayed.
Tare
With a tare (container) placed, the current mass value is set to zero, and the net is
displayed. This is the same as the
TARE
key.
Template T
Command
Reply
The current mass value becomes zero and the net mark is displayed.
Cancel of tare
The display value becomes the gross and the net mark is turned off. (The tare value
becomes zero.)
Template CT
Command
Reply
Gross is displayed and the net mark is turned off.