48
If the command is S
,
T
or
Z , the scale will send a second acknowledgment
<ACK><C
R
><L
F
> or <ACK> (see F-Function “
f-09-02
”) when the command
operation is completed.
If an error occurs, the scale will send an error code.
The error format is
E C , E n C L
F
, “n” being error number.
En Definition
Notes
E0
Communication Error
Parity error, framing error, etc.
E1
Undefined command Error
Command does not exist for the scale.
E2
Scale not ready Error.
The scale is not in a state where a command
could be expected.
E4
Too many characters Error
Command contains too many characters.
E6
Format Error
Command contains invalid characters.
E7
Out of range Error
Value is out of range. Tare weight is more than
the capacity, etc.
Data Format
Examples below are for
f-09-02
=”
0000
”. <ACK>=06H.
Store unit weight and other value in use (according to
f-01-05
).
Command E , 1 2
C L
F
Stores to
id-12
. (E,000012C
R
L
F
is acceptable.)
Reply ACK C L
F
ID Number
Command ? I D
C L
F
Reply
I D , 0 0 0 0 1 2 C L
F
PCS (Count ) Data
Command ?
Q T
C L
F
Reply
Q T
,
+ 0 0 0 0 1 2 3 4
[ P C C L
F
Stable Positive Data
U S
,
-
0 0 0 0 5 6 7 8
[ P C C L
F
Unstable Negative Data
O L
,
+ 9 9 9 9 9 9 9 9
[ P C C L
F
‘E’ display
Weight Data
Command ?
W T
C L
F
Reply
S T
,
+ 0 0 1
.
2 3 4 6
[ k g C L
F
Stable Positive Data
S T
,
-
0 0 2
.
7 2 5 5
[
l
b C L
F
Stable Negative Data
U S
,
-
0 0 1 2
.
3 4 6
[
l
b C L
F
Unstable Negative Data
U S
,
+ 0 0 0 5
.
5 9 3
[ k g C L
F
Unstable Positive Data
O L
,
+ 9 9 9 9
.
9 9 9
[ k g C L
F
‘E’ display
O L
,
-
9 9 9 9
.
9 9 9
[
l
b C L
F
‘-E’ display
“
[
” in examples below shows “Space” (20H).