35
F-06-X RS-232C Data Output
F-06-X requires OP-03 or OP-04 RS-232C interface.
Data Out Mode
0
Key Mode: Data is sent by pressing the PRINT key.
+ Command Mode.
1
Stream Mode: Data is sent continuously.
Command Mode cannot be used.
2
Auto-Print Mode A: Data is sent if the weight display is stable
at +5d (weighing display division) and above.
+ Command Mode.
3
Auto-Print Mode B: Data is sent if the weight display is
stable, at ±5d (weighing display division) and above/below.
+ Command Mode
4
Command Mode Only.
5
To use as Bar Code Reader Interface.
6
UFC format with Key Mode (see setting “
0
”).
7
UFC format with Auto-Print Mode A (see setting “
2
”).
8
UFC format with Auto-Print Mode B (see setting “
3
”).
Data to be Sent.
0100
Count data sent.
You select which data to be sent by keying in a
0 or 1
for the data: ID no., PCS (count),
weight or unit weight.
Example:
Key in 1
1
0
0 to display
1100
,
this setting would send only the ID number and
the count.
Data Format
0
Format for AD-8121 MODE 1.
1
Format for AD-8121 MODE 3.
No difference between “
0
”
and ”
1
” when used with the
UFC format.
2
Format for general apparatuses, computers, etc.
Baud Rate
0
2400 bps.
1
4800 bps.
2
9600 bps.
0000
Weight
PCS(count)
Unit
ID Weight
f-06-02
f-06-04
f-06-01
f-06-03