57
Transmitting all memory data at one time
Confirm that the “Serial interface (
5if
)” parameters are set properly. For details, refer to “10.
FUNCTION TABLE” and “17. CONNECTION TO PERIPHERAL EQUIPMENT”.
1 Press and hold the PRINT key until
reCall
is displayed.
2 Press the SAMPLE key to display
out
.
3 Press the PRINT key to display
Sout no
with “
no
” blinking.
5 Press the RE-ZERO key to display
out go
with “
go
” blinking.
6 Press the PRINT key to transmit all data using the RS-232C interface.
7 The balance displays
Clear
when all data is transmitted. Press the CAL key to return to the
weighing mode.
Deleting all memory data at one time
1 Press and hold the PRINT key until
reCall
is displayed.
2 Press the SAMPLE key several times to display
Clear
.
3 Press the PRINT key to display
SClr no
with “
no
” blinking.
4 Press the RE-ZERO key to display
SClr go
with “
go
” blinking.
5 Press the PRINT key to delete all data
6 The balance displays
end
, then
reCall
when all data is deleted.
7 Press the CAL key to return to the weighing mode.