10
7.
Installation
1. Adjust the horizontal level of the product using the
four leveling feet and bubble spirit level. Confirm that
the scale is seated firmly and does not rattle (refer to
“
4.1. Precautions When Installing the Scale
”).
Perform sensitivity adjustment when the scale is installed for the first time or relocated (refer to
”).
Ground the scale by using the grounding terminal when it is used where static electricity occurs.
2. Connect the AC adapter to an outlet, and the USB cable to the power input terminal that is on the side of
display.
Performance is not guaranteed if a commercial AC adapter or a USB cable for other products is used.
When using a mobile battery as a power source, it is recommended that you use one where power
can be turned on/off with a switch.
Set function setting
to “ ” if the mobile battery has an automatic power-off function
(Operating time will be shortened) (refer to “
”).
The table below shows the reference values of continuously usable time with mobile batteries
(comparator function being off, no connected devices).
Continuously usable time
5000 mAh
10000 mAh
Approximately 24 hours
Approximately 50 hours
Performance is not guaranteed with all mobile batteries.
Confirm the specifications of mobile batteries before use. Product damage due to mobile batteries
will void the warranty.
Please confirm that the local voltage and the receptacle type are correct for your scale.
Leveling foot
Bubble spirit level
USB cable
AC adapter
Outlet
Power input terminal
Grounding terminal
The scale is horizontal.
The scale is tilted.
Weighing will be correct
Weighing will not be correct