Ver.D20191212B
1
N
ADENDUM to FX-iN/FX-iWPN Legal For trade
(
LFT
)
Approved
Models
Purchased FX-iN/FX-iWPN balances are LFT/Legal for Trade. These balances can be set following the
regulations for the US and Canadian markets. However, there are some differences and limitations in
comparison to the standard versions FX-i/FX-iWP balances, as detailed below.
Note: It is fully the responsibility of the purchaser to set up the balances, including calibration and sealing,
in accordance with local weights and measures regulations.
1.
Display and key Operation
General Key operation
For the
CAL, MODE, SAMPLE,
and
keys, please refer to “4. DISPLAY SYMBOLS
AND KEY OPERATION” of the standard instruction manual.
Key Operation for LFT Approved Model
①
The
ON:OFF key
becomes the
ON:OFF/ZERO key
When this key is pressed and hold for 2 seconds: The display turns ON/OFF.
When this key is pressed once: The display is zeroed out.
For the other functions of the
ON:OFF
key mentioned in the general instruction manual, please
use the
ON:OFF /ZERO
key as the equivalent of the
ON:OFF
keys in the standard versions.
②
The
RE-ZERO
key in the standard version becomes the
TARE
key on the legal for
trade version.
When this key is pressed: The Tare value is registered/canceled.
For the other functions of the
RE-ZERO
key mentioned in the instruction manual,
please use the
TARE
key.
③
Front label
The front label must be changed in accordance with the selected mode.
(Example: FX120iN)
This is purchasers’ responsibility
or
①
②
③
Default label (e=10d)
Label in the products’ box (e=d)
This document hosted by: www.oldwillknottscales.com
Содержание FX-iWP Series
Страница 5: ...Ver D20191212B 5 _ Space ...
Страница 12: ...5 ...
Страница 50: ...43 9 7 Data Format Examples ...
Страница 79: ...72 MEMO ...