Step 1    Pass the cable B through the pole.
Step 2    While taking care not to pinch the cable, secure the pole to the weighing 

   platform, using two screws.

This manual describes how the EM Series Digital Platform Scales work.

Step 3    Secure the bracket to the pole, using a nut and screw.

Please read this manual completely before using the scale.

Step 4    Connect cable A to cable B.
Step 5    Bundle the excess cable and insert it into the pole.
Step 6    Attach the display to the bracket.

CAUTION

   

 

The scale is a precision instrument. When installing the scale, consider the

     following conditions to get the most out of your scale.
        Install the scale where it is not exposed to direct sunlight, where the temperature
        and relative humidity is stable and where there is no dust and draft.
        Install the scale on a solid and level surface which is free from vibration.
     The scale is not waterproof. Do not use it where it will get wet.
     Do not disassemble the scale. Do not open the case to repair or modify it. That 
     may cause a fire or damage to the scale. Damage caused under such conditions 
     will void the warranty.
     If the scale malfunctions, stop using it immediately. Contact the local A&D dealer.

    Three weight display resolutions are available, normal (1/3000), high (1/6000 or 
    1/7500), and higher (1/12000 or 1/15000).
    Two weighing platform sizes are available. The EM-KAM has a smaller pan and 
    the EM-KAL has a larger pan.
    As a power source, either the AC adapter or internal rechargeable battery can be used.
    The counting function easily counts the number of objects of the same weight.
    The optional EM-03 serial interface can transmit the weight value to a personal
     computer or a printer.

The following options are available for the EM series:
     EM-03     RS-232C serial interface
     EM-12     Adjustable bracket
     EM-13     Stabilizing foot ( for EM-30KAM , -60KAM ) 
     EM-14     Desktop / Wall-mount stand

When unpacking, check whether all of the following items are included.

     EM-15     Protective cover

The EM series scale 

screws and nut  

AC adapter  

Instruction manual

     EM-16/17     Extension cable ( 3m/5m )

Note:
Confirm the AC  adapter type
is correct for your local voltage

     Adjust the level of the scale, using the leveling feet. Confirm that the bubble of

and receptacle type.

     the spirit level is in the center of the level.
     The best installation conditions are: stable temperature (-10

 to 40

℃)

 and

     relative humidity ( less than 85% ), solid and level surface, no drafts or vibration, 
     no direct sunlight.
     Do not install the scale where there may be static electricity or sudden changes
     in temperature.

EM series (Common for all models)

Step 1 Press the 

ON/OFF

 key to turn the power ON.

The display shows “ - - - - - ” if there is something placed on the weighing
pan or the scale is unstable.

Step 2 Press the 

UNITS/MODE

 key to select “kg” ( kg mode ).

Step 3 When the display doesn't show zero, press the 

RE-ZERO

 key.

Step 4 When a tare ( container ) is used, place the container on the weighing pan

and press the 

RE-ZERO

 key to set the display to zero.

Scale weighing range = weighing capacity - tare weight

Step 5 Place the object to be weighed on the weighing pan or in the container. Wait

for the STABLE indicator to turn on and read the value.

Step 6 Remove the objects from the weighing pan.
Step 7 Press the 

ON/OFF

 key to turn the power OFF.

When the auto power-off function is enabled, the scale turns off automatically
after 5 minutes of stable condition and Error displayed.

Display and symbols

Step 1 Press the 

UNITS/MODE

 key to select “pcs”.

Step 2 Press and hold the 

UNITS/MODE

 key to enter

the sample unit weight storing mode.
( The figure on the left is the number of samples. )

Step 3 To change the number of samples, press the

ON/OFF

 key. It may be set to 5, 10, 20, 50 or 100.

Step 4 If necessary, place a container on the weighing 

pan and press the 

RE-ZERO

 key. Confirm that

the right side of the number of samples shows zero.

Step 5 Place the correct number of samples on the

weighing pan or in the container. In this example,
place 20 pieces of samples.

Step 6 Confirm that the STABLE indicator is turned on.

Press the 

UNITS/MODE

 key to calculate and

store the unit weight.
The scale is set to count object with this unit weight.
The unit weight is stored in memory, even if the
power is turned off.

Note

     The total weight of samples should be more than shown below, regardless of the
     number of samples.
     Weighing capacity of 30 kg :100g, capacity of 60 kg :200g, capacity of 150 kg:500g
     If not, the display shows “ 

lo ut

 ” and returns to the display of step 5. Increase the

     number of samples (step 3) and try again.

Turns on when the weight value is stable.

Turns on when the scale shows zero.

 STABLE

 ZERO

 NET

Turns on when the NET weight is displayed.
( The tare operation is in progress. )

 HOLD

Turns on when the scale hold the weight.

 Battery indicator

Changes as the battery capacity decreases as
shown to the left.

 Weighing units

“kg”

Basic weighing mode in kg.

“pcs” (pieces)

Counting mode which determines

a unit weight (the weight of one piece) from the
known pieces of samples and calculates how
many pieces are on the weighing pan using the
unit weight.

/

1WMPD4002222C

1.1.  Precautions

1.2.  Features

1.3.  Unpacking

1

. 

INTRODUCTION

EM SERIES

Digital Platform Scale

3

. 

INSTALLATION

3.1.  Setting up the Scale

Screw

Step 2

Cable A

Bracket

Cable B

Step 4

Step 1

Step 5

Step 3

Step 6

Display

Nut

Pole

Leveling foot

3.2.  Installing the Scale

2.  DESCRIPTION OF EACH PART

Display

Leveling foot

Weighing pan

Spirit level

Pole

AC adapter jack

Calibration switch

Battery indicator

STABLE indicator

NET indicator

ZERO indicator

ON/OFF key

RE-ZERO key

Weighing units

UNITS / MODE key

4

. 

BASIC WEIGHING OPERATION

5

. 

COUNTING MODE

     0

pcs

   5 0

pcs

Number of samples

  20 0

pcs

Confirm zero

  20 -

pcs

    20

pcs

Container

Weighing pan

Spirit level

Screws

HOLD indicator

EM-03 option

RS232C 

D-sub 9 pin