6. Calibration
This function adjusts the balance for accurate weighing.
Perform calibration in the following cases.
□
When the balance is first used.
□
When the balance has been moved.
□
When the ambient environment has changed.
□
For regular calibration.
6-1 Calibration using a weight
□
Prepare a calibration weight (optional) before start.
1. Warm up the balance for at least half an hour with nothing
on the pan.
2. Press and hold the calibration (CAL) switch until
appears, and release the switch.
3. The balance displays
.
To change the calibration weight value, proceed to step 4.
To use the calibration weight value in the balance memory,
proceed to step 5.
4. Press the
key. The display shows the
calibration weight value in “gram” that is stored in the
balance. Use the following keys to change the value.
key
To select the digit blinking to change.
key To set the value of the digit selected.
key
To store the value and return to
step 3.
key
To cancel the value and return to
step 3.
□
Use a calibration weight of more than 2/3 of the
capacity (of the highest range for EW-
i
series).
Calibration
(CAL) switch
Press and lower the
calibration switch cover.
Press and hold
the CAL switch.
Release
the CAL switch.
Set the weight using
the relevant keys.
Содержание EK-i Series
Страница 49: ...CT 49 3 ON OFF 100 EW i pcs 100 ACAI OP 04 09 RS 232C...
Страница 50: ...CT 50 4 4 1 1 2 3 6 AC ON 30 6 4 2 EKW 09i EK i EW i EKW 09i EK i EW i...
Страница 51: ...CT 51 5 5 1 ON OFF 1 ON ZERO 10 ON NET ZERO 2 OFF...
Страница 52: ...CT 52 6 6 1 1 30 2 CAL 3 4 5 4 3 3 2 3 EW i CAL CAL CAL...
Страница 53: ...CT 53 5 3 6 7 CAL...
Страница 54: ...CT 54 7 7 1 A D 7 2 100 100 100 ACAI CAL OP 09...
Страница 55: ...CT 55 OFF ON A D...
Страница 57: ......