Daily check
Display settings: MENU key
Calibration test/check button
Daily check button
[Daily check] screen, [Environmental conditions] screen (From this screen, display
check items in sequence)
Daily check details
During daily check, you check the minimum required items for using the balance to make sure that there
is no serious error with the balance. It is advisable to perform this check every day before work, though
this may depend on the maintenance level of the balance.
1 Environmental conditions
Make sure that the operating temperature and humidity of the balance fall within the specifications.
If the humidity is 45 %RH or less, it is recommended that static elimination be performed before
weighing.
Be sure to warm up the balance before use for at least an hour, or at least four hours for BA-6TE/6DTE,
with the AC adapter connected to the power supply.
The warm-up time is the period of time when the power to the balance is on before use.
*
You can measure up to 9999 hours for the warm-up time.
1
Environmental conditions
2
Check item position
4
Next button
3
Back button
Name
Description
1
Environmental
conditions
Displays the environmental conditions.
2
Check item position Displays the position of the current check item.
3
Back button
Returns to the [Calibration test/check] screen.
4
Next button
Considers this check item as OK and proceeds to the next item.
Содержание BA-T Series
Страница 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Страница 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Страница 107: ...107 Output example for daily check output results Daily check output results Daily check PDF results...
Страница 117: ...117 Output example for periodic check output results Periodic check output results...
Страница 118: ...118 Periodic check results PDF...
Страница 129: ...129 AND MEET analysis graph example...
Страница 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...
Страница 219: ......