AD-8920A REMOTE DISPLAY INSTRUCTION MANUAL
㸦᪥ᮏㄒࡢྲྀᢅㄝ᫂᭩ࡣᑐ㠃ࢆࡈぴࡃࡔࡉ࠸ࠋ㸧
1. Introduction
The AD-8920A is a remote display for displaying the weighing data transmitted by an A&D
manufactured weighing instrument, using either RS-232C or current loop.
Applicable weighing instruments (electronic balances/platform scales) are listed below.
What the package contains
Table 1 Applicable instruments, required options and cables (As of October, 2008)
࣭
AD-8920A main unit
1 unit
Using RS-232C
Using current loop
࣭
AC adapter
1 pc
a
Weighing
instrument
Option
required
Communication
cable
Option
required
Communication
cable
a
AD-4212A/B
None
D-SUB25 cable
Not applicable
a
EK-H, EK-HR
OP-03H
D-SUB25 cable OP-05H
DIN cable
࣭
Identification label
2 pcs
EK-i, EW-i
None
D-SUB9 cable
Not applicable
(One each for AC adapter and cable) ET-W,
ET-WR
OP-03W
D-SUB25 cable OP-05W
DIN cable
࣭
Instruction manual (this leaflet) 1
copy
FC-i
None
D-SUB9 cable
Not applicable
࣭
Communication cable (Approx. 1 m)
1 pc
FG
OP-03
DIN cable
OP-05
DIN cable
When ordering, a cable appropriate to the
FG-M/L
OP-23
DIN cable
Not applicable
instrument connector type is included, selecting FS
OP-03
DIN cable
Not applicable
one from below. (Refer to Table 1 for details.)
D-SUB9 cable / D-SUB25 cable / DIN cable
FX-i, FX-CT, FX-GD,
FXi-WP, FZ-i
None D-SUB9
cable
Not applicable
GH, HR-300i/202i
None
D-SUB9 cable
Not applicable
Accessories (Sold separately)
GR
None
D-SUB25 cable
Not applicable
࣭
Communication cable (Approx. 5 m)
D-SUB9 cable
AX-KO3412-05M
D-SUB25 cable
AX-KO1864-05M
GX, GX-R, GF,
GX-K, GX-KR,
GF-K, GP, GP-R
None D-SUB25
cable
OP-04
or
OP-06
DIN cable
DIN cable
AX-KO3413-05M
HC-i
OP-03
DIN cable
Not applicable
࣭
Communication cable (Approx. 10 m)
D-SUB9 cable
AX-KO3412-10M
HR-200/120/60 OP-03
D-SUB25
cable
OP-03 or
OP-05
DIN cable
D-SUB25 cable
AX-KO1864-10M
HV-G/HW-G
None
DIN cable
Not applicable
DIN cable
AX-KO3413-10M
Connection diagram
2. Preparation
2-1. Setting the weighing instrument
(1) Set the data output mode to “Stream mode”.
(2) For the other settings, refer to Table 2.
Table 2 List of instrument settings
Item Setting
Description
Data output mode
Stream mode
Outputs the weighing data continuously.
Baud rate
600, 1200, 2400, 4800, 9600 bps
AD-8920A recognizes the baud rate automatically.
Length, Parity bit
7 bits-even, 7 bits-odd, or 8 bits-none
AD-8920A functions correctly with any one of those
Stop bits
1 bit or 2 bits
AD-8920A functions correctly with either one.
Terminator
<CR> or <CR><LF>
AD-8920A functions correctly with either one.
Data format
A&D standard format
CTS control
No control of CTS, RTS
Output (hardware)
RS-232C or current loop
AD-8920A recognizes the output mode automatically.
Note) The available items depend on the weighing instrument and may not be available when the settings are fixed.
For a detailed description of the settings, refer to the instruction manual for the weighing instrument used.
2-2. Connecting the AD-8920A
(1) Refer to Table 1 shown above to confirm that the communication cable is of the correct type.
(2) Disconnect the AC adapter from both the weighing instrument and the AD-8920A.
Refer to the connection diagram shown above to connect the AD-8920A to the weighing
instrument, using the specified communication cable.
CAUTION
The D-SUB9 cable (standard
or accessory) is for use only
with the AD-8920A.
If used with an older AD-8920,
it could blow the fuse or cause
other damage to the weighing
instrument.
Attach the identification label
on the cable so that it will be
used only with the AD-8920A.