Adjust the desire flow output value using the joystick
UP
,
DN
,
LEFT
and
RIGHT
buttons. Press the joystick
ENT
button to activate analog output.
To abort the analog output test mode, press the
Esc
button. Once the test
mode is deactivated, the analog output should represent actual flow rate
readings.
6..4
4..113
3..7 S
Stta
attu
us
s L
LE
ED S
Se
ettttiin
ng
gs
s
DPM series Mass Flow Meters are equipped with dual color LED which allows
signaling a variety of different events with combinations of three colors (red,
green and amber) and a specific time pattern. Status LED operation can be
adjusted/ filtered for the indication of different events based on custom user
needs.
Status LED can be set to the following modes (
1. Normal, which supports the following events:
1.1 Auto Zero Failure (constant RED)
1.2 Fatal Error (constant RED, requiring the system to be reset for recovery)
1.3 User entry via side Push Button (specific pattern limited by a time interval
up to 35 seconds)
1.4 Power Up Sensor Warm Up interval (1 to 3 seconds). (Constant
AMBER). This can be interrupted only by User PB entry or Fatal Error.
2. Monitoring Flow Alarm and Flow Totalizer events (default settings):
2.1 High Flow Alarm RED/OFF (alternating every second)
2.2 Low Flow Alarm GREEN/OFF (alternating every second)
2.3 Totalizer#1 Event AMBER/OFF (alternating every second)
2.4 Totalizer#2 Event AMBER/OFF (alternating every second)
2.5 High Flow Alarm and Totalizer#1 Event RED/AMBER (alternating every
second)
2.6 High Flow Alarm and Totalizer#2 Event RED/AMBER (alternating every
2 seconds)
2.7 Low Flow Alarm and Totalizer#1 Event GREEN/AMBER (alternating
every second)
2.8 Low Flow Alarm and Totalizer#2 Event GREEN/AMBER (alternating
every 2 seconds)
2.9 Both Totalizer#1 and Totalizer#2 Events AMBER/OFF (on for 3 seconds,
off for 1 second)
2.10 High Flow Alarm and Totalizer#1 & #2 Events AMBER/RED (AMBER
for 3 seconds, RED for 1 second)
2.11 Low Flow Alarm and Totalizer#1 & #2 Events AMBER/GREEN (AMBER
for 3 seconds, GREEN for 1 second)
CAUTION:
When “Analog Output Test” is selected, the output
reading does not represent any actual Process Information (PI)
variable (flow rate reading).