When the DPM device is set as the last device on the long RS-485 bus segment,
the 120
ȍ
bus termination resistor must be connected between the RS-485 (+)
and (-) terminals close (< 6 feet) to this device.
NOTE:
The DPM instrument offers an integrated switchable 120
ȍ
W
ermination resistor between the RS-485 (+) and (-) pins. On
instruments with a local display and joystick interface, the 120
ȍ
termination resistor can be activated (enabled) using General
Settings / Communication Port / RS-485 Termination menu selection.
By default, the instrument is shipped from the factory with the RS-485
Termination mode set to Disabled.