10.4 Recommended separation distances between portable and mobile RF
telecommunications equipment and the laser
The laser is intended for use in an electromagnetic environment in which the RF disturbances are controlled. The customer or the
user of the laser can help to avoid electromagnetic interference by maintaining the minimum distance between portable and
mobile HF telecommunication devices (transmitters) and the laser - depending on the output power of the communication device,
as stated below.
Test
frequency
MHz
Frequency
band
MHz
Radio
service
Modulation
Maximum
performance
W
Distance
m
Immunity test level
V/m
385
380 to 390
TETRA 400
Pulse
modulation
18Hz
1.8
0.3
27
450
430 to 470
GMRS 460
FRS 460
FM
± 5 kHz Hub
1 kHz Sinus
2
0.3
28
710
745
780
704 to 787
LTE Band
13,17
Pulse
modulation
217 Hz
0.2
0.3
9
810
870
930
800 to 960
GSM
800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation
18 Hz
2
0.3
28
1720
1845
1970
1700 to
1990
GSM 1800,
CDMA 1900,
GSM 1900,
DECT,
LTE Band
1,3,4,25,
UMTS
Pulse
modulation
217 Hz
2
0.3
28
2450
2400 to
2570
Bluetooth,
WLAN
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse
modulation
217 Hz
2
0.3
28
5240
5500
5785
5100 to
5800
WLAN
801.11 a/n
Pulse
modulation
217 Hz
0.2
0.3
9