3M Prevena 125 Скачать руководство пользователя страница 1

This is a quick guide, further details are in the 3M

 Prevena

 Therapy Clinician Guide

Alert type

How to fix

Leak Alert

One beep, one solid  

yellow light

To ensure proper application of the 3M

 Prevena

 Incision Dressing, the Prevena

 

125 Therapy Unit 

provides the Visicheck

 Feature. Double press ON/OFF button. 

- One light: Best 

- Two lights: Good 

- Three lights: Marginal 

If the Visicheck

 Feature indicates a “Marginal” leak rate condition, follow these steps: 

- Ensure canister is locked into the unit

- Check the tubing connectors to make sure they are fully engaged and locked

- Use finger to smooth out any creases in the adhesive

- Use Prevena Incision Dressing Patch Strips (located in the dressing package) to help seal leaks. 

- Slowly press firmly around dressing edges 

Canister Full Alert

Two beeps, one solid 
yellow light

Check fluid level in the canister
1.  If the canister isn’t full check for kinks or pinched tubing

2.  If the canister is full then turn the unit off

3.  Call 1-800-275-4524

Low Battery Alert

Low- One slow beep, one 

solid yellow light

Critical- One beep, 

repeating rapidly, 

increasing volume, two 

solid yellow lights

This therapy unit comes with three AA batteries
1.  Hold five second to turn off

2.  Press and push the panel on the back of the unit to open

3.  Insert three AA batteries (lithium batteries recommended)

4.  Press and push to close panel

5.  Hold two seconds to turn on

System Error Alert

One Beep, repeating 

rapidly, increasing in 

volume, two solid yellow 

lights

1.  Hold for five seconds to turn off

2.  Hold two seconds to turn back on

3.  If the therapy is still not running call 1-800-275-4524

Device Lifecycle 

Expired

Three lights show life 

cycle is up. There will be 

a repeated beep for 15 

seconds and then the unit 

will shut off.

Call 1-800-275-4524

Life cycle=192 hours (eight days)

3M

 Prevena

 125 Therapy Unit 

indicators and alerts

Caution 

Indicator

Audible Alarm 

(muted)

Battery 

Level

Quick Guide

Incision Therapy