Inserting and removing 3M
™
Clean-Trace
™
test swab into the LM1
Luminometer
1.
Open the 3M Clean-Trace LM1 Luminometer by pressing the rocker cap.
2.
Insert an activated 3M Clean-Trace test swab into the chamber, making sure
the cuvette is at the bottom of the chamber.
3.
The 3M Clean-Trace Test handle should be visible at the top of the chamber.
4.
Release the rocker cap. The user may now start the ATP test run for the 3M
Clean-Trace Test swab inside the chamber.
5.
After the ATP test is completed, open the chamber by pressing the rocker cap.
6.
To remove the 3M Clean-Trace Test, grasp the top of the test swab and pull
upwards. Follow the instructions for disposal in the 3M Clean-Trace Test
Instructions for Use. Consult with local regulations for any additional
regulations regarding disposal.
c.
Using the Base
For extra stability and convenience, the 3M Clean-Trace LM1 Luminometer can be
placed on the 3M™ Clean
-
Trace™ Base. The 3M Clean
-Trace Base holds the
Luminometer at an optimum working and viewing angle.
The 3M Clean-Trace LM1 Luminometer also has a fold-out stand built into the
bottom of the instrument. The fold-out stand allows the instrument to stand
without the base.
d.
Battery usage and maintenance
▪
The 3M Clean-Trace Luminometer holds an internal, rechargeable lithium-ion battery that must
be charged to use the 3M Clean-Trace Luminometer.
▪
In normal usage, the battery provides up to eight hours of operation from full charge.
▪
As the battery ages, it will provide less than eight hours of operation from full charge.
▪
The battery must be charged when the indicator is red, indicating less than 10% charge capacity.
▪
The battery does not need to be fully discharged before recharging.
▪
When storing a 3M Clean-Trace Luminometer for more than a month, the battery level should be at minimum
of 50% charge. Do not store a 3M Clean-Trace Luminometer with a fully discharged battery, it may cause the
battery to lose the ability to be recharged.
▪
Battery replacement is recommended after two years of normal use. For battery replacement, the 3M Clean-
Trace Luminometer must be sent to a 3M Service Center.
e.
Charging the battery
▪
The 3M Clean-Trace LM1 Luminometer is shipped with a USB cord, a charger and various charger power
adapters.
▪
Select the charger adapter that is appropriate for power outlet in your country.
▪
Connect the charger to the 3M Clean-Trace Luminometer using the provided USB cord.
▪
Connect the charger to a power outlet.
▪
The battery will start charging and becomes fully charged in approximately 2 hours.
▪
To monitor the charge status, turn on the 3M Clean-Trace Luminometer. A lightning bolt icon appears in the
battery charge level. Refers to the
Battery level indicator
section for more details.