2
EN
(English)
III.
ENVIRONMENTAL OPERATING CONDITIONS
Environmental Condition
Operating Condition
Indoor use only
▪
Only connect the data connections to a computer that
conforms to IEC/EN/UL/CSA 60950-1
▪
Keep 3M
™
Clean-Trace
™
Luminometer away from liquids
and high humidity when connected to the computer
Altitude
3000 meters maximum
Operating temperature rRange
+5 to +40 , (+41 to +104) °C (°F)
Relative humidity
10 - 90 Non-condensing %RH
Storage temperature range
-40 to +70 (-40 to +158) °C (°F)
AC Voltage for power supply
Mains supply voltage fluctuations up to + 10% of nominal
voltage
Temporary overvoltage occurring on the mains supply
100 -240 Volts
Frequency
50-60 Hertz
AC Current for power supply
300 mA
Power supply output voltage
5 Volts DC
Power supply output current
1.2 Amps
Luminometer input current
850 mAmps
Overvoltage
Category II
Pollution degree
2
IV.
INSTRUMENT SPECIFICATIONS
Specification
Length
59.27 (2.33) mm (in)
Width
88.47 (3.48) mm (in)
Height
268.17 (10.56) mm (in)
Weight
517.83 (18.26) g (oz.)
Physical connector
USB 2.0 - Type mini
Wireless connectors
Wi-Fi 2.4 GHz & 5GHz
Bluetooth® 2.0