Performing an ATP Test run
Follow the Instructions for Use for the
3M™
Clean-T
race™
Tests swabs (UXL100 or AQF100). Collect a sample, activate
the test swab and shake it for at least 5 seconds.
▪
Open the 3M
™
Clean-Trace
™
Luminometer by pressing the rocker cap. Then insert the activated test into the
chamber. Keep the 3M Clean-Trace Luminometer upright whenever a test is in the chamber to ensure accurate
results and avoid spillage of reagents, which could damage the sensitive components of
the instrument.
▪
Tap
[Start]
to begin measuring. A progress bar appears and moves from left to right.
▪
The test result appears and is stored in the memory so it can be synchronized with the 3M
Clean-Trace Hygiene Management Desktop Software. The result is also interpreted using
your established Pass/Fail criteria and is displayed on the screen as
Pass
,
Caution
, or
Fail
.
▪
Remove the Test swab from the 3M
™
Clean-Trace
™
Luminometer; press the rocker cap, grasp the top of the Test swab,
and pull upwards. Follow current industry standards and local regulations for disposal of used Test swabs.
o
To test another Test point, tap the next Test point at the bottom right of the LM1 screen, or Previous Test point at
the bottom left of the 3M Clean-Trace LM1 screen.
o
To do a retest, tap
[Retest]
then tap
[Start].
Retest is allowed up to 12 hours after the initial test.
o
To display the list of Test points, tap
[List View].
If result is within the Fail criteria, the
screens below will be presented
If result is within the pass criteria, the
screens below will be presented
BLENDER
If result is between the Pass and the Fail
limits, the screens below will be presented