64220, Rev. 1/April 2019
3M™ GMI™ PS200-LMP AUTO BUMP/CALIBRATION STATION
SPECIFICATIONS
A-1
APPENDIX A. SPECIFICATIONS
Table A-1: Windows Compatibility
PS200-LMP settings are tested to run on the
following Windows operating system:
Windows 7, Windows 10
Table A-2: Gas Cylinder Permissible Concentrations
Gas 1, % Methane (CH
4
):
1.0% to 4.0%
Gas 1, PPM Carbon Monoxide (CO):
0ppm to 1000ppm
Gas 2, % Methane (CH
4
):
80% to 100%
Table A-3: Calibration Parameters
Gas 1 Application Time:
60 seconds
Gas 2 Application Time:
30 seconds
Total Calibration Time:
3 minutes, 30 seconds
Gas 1, % Methane (CH
4
), Pass Limits:
±0.2% (of target)
Gas 1, PPM CO, Pass Limits:
±4% (of target)
Gas 2, % Methane (CH
4
), Pass Limits:
±3% (of target)