Cause

•  Solution

The LED showing the 

operating status in the charger 

flashes alternately red and 

green. The handpiece is in the 

charger.

Malfunction during charging.
•  Remove and re-insert the 

battery to reset the unit.

•  The battery is damaged.
•  Replace the battery.

The LED showing the 

operating status in the 

charger does not come  

on although the plug has 

been connected to an 

electrical outlet.

Power outlet carries no 

voltage.
•  Use a different power 

outlet. 

Charger is defective.
•  Have the charger 

repaired.

The light intensity is too low.

•  Clean the light guide 

and the protecting 

glass in the light guide 

mounting hole (please 

refer to “Cleaning the 

Light Guide”).

The LED showing the 

operating status in the charger 

flashes alternately red and 

green. The handpiece is not in 

the charger.

Charger is defective.
•  Unplug and plug in the 

charger to reset.

•  Have the charger 

repaired.

During light emission in 

extended mode, an error 

signal sounds for 2 seconds,  

the emission is stopped,  

and the handpiece changes 

to sleep mode.

The handpiece has become 

overheated in the course 

of use. The handpiece can 

be used again once it has 

cooled down.
•  Allow the handpiece to 

cool down.

-   Immediately after the 

temperature control 

has turned off the 

handpiece, it must cool 

down a minimum of 1.5 

minutes in the air before 

a new light emission can 

be started.

-   The handpiece can 

be used again, once 

pressing the START 

button successfully 

activates the light.

-   During the cooling-off 

period immediately after 

the acoustical signal 

has gone off, 2 minute 

cooling-off time in the air 

typically corresponds to 

1 minute operating time.

The light emission does not 

start when the START button 

is pressed; an error signal 

sounds for 2 seconds.

The handpiece has become 

overheated in the course 

of use. The handpiece can 

be used again once it has 

cooled down briefly.
•  Allow the handpiece to 

cool down.

-   The handpiece can 

be used again, once 

pressing the START 

button successfully 

activates the light.

-   During the cooling-off 

period immediately after 

the acoustical signal 

has gone off, 2 minute 

cooling-off time in the air 

typically corresponds to 

1 minute operating time. 

The power level display of 

the handpiece flashes red 

while the handpiece is in  

the charger.

Charging problem. The 

battery is defective or at the 

end of its useful life.
•  Remove and re-insert the 

battery to reset the unit.

•  Replace the battery.

The handpiece has not been 

used for a long time and 

now it cannot be turned on.

There is not enough charge 

in the battery to turn on the 

handpiece.
•  Place the handpiece  

in the charger and  

re-charge the battery.

The power level display of 

the handpiece flashes red.
The ongoing exposure is 

interrupted (light off signal 

is emitted) followed by 

a 2 second-error signal; 

the handpiece switches to 

“sleep” mode and resists 

further activation.

The battery lacks sufficient 

charge.
•  Place the handpiece  

in the charger and  

re-charge the battery.

The charge level display  

on the handpiece glows  

red steadily. 

The remaining battery 

charge is adequate only 

for typically 20 3-second 

exposure cycles.
•  Place the handpiece  

in the charger and  

re-charge the battery.

Error

Cause

•  Solution

Troubleshooting

Содержание Ortholux Luminous Curing Ligh

Страница 1: ...ENGLISH Fast Curing Cordless LED Light Instructions For Use Ortholux Luminous Curing Light...

Страница 2: ...age stated on the rating plate is compatible with the available main voltage Operation of the unit at a different voltage may damage the unit 3 Position the unit so that the power plug is accessible a...

Страница 3: ...devices 18 Do not use Ortholux Luminous Curing Light for patients with a history of photobiological reactions including individuals with urticaria solaris or erythropoetic protoporphyria or who are c...

Страница 4: ...The handpiece switches to sleep mode once it is placed in the charger or if left unused for approximately 5 minutes outside the charger NOTE Do not discard these Instructions for the duration of produ...

Страница 5: ...o technical modification without prior notice Installation of the Unit Factory Settings The factory settings of the unit are as follows When first turned on the unit is in Extended Mode i e only the P...

Страница 6: ...ge the new battery completely for the first time The status LED showing the operating status of the charger flashes green during the charging operation see below Operating Status Display of the Charge...

Страница 7: ...om the Extended Mode Caution High intensity light always involves the production of heat When utilizing the light in Extended Mode 15s do not keep the light in one position Longer exposure in the regi...

Страница 8: ...varying arrangement of the components of the units NOTE The light intensity is independent of the battery charge level Caution Clean the testing area with a damp cloth Without applying pressure place...

Страница 9: ...y after the acoustical signal has gone off 2 minute cooling off time in the air typically corresponds to 1 minute operating time The light emission does not start when the START button is pressed an e...

Страница 10: ...defective batteries and units in accordance with local legal regulations Maintenance and Care The Ortholux Luminous Curing Light is maintenance free with the exception of cleaning the light guide and...

Страница 11: ...A PARTICULAR PURPOSE User is responsible for determining the suitability of the product for user s application If this product is defective within the warranty period your exclusive remedy and 3M Unit...

Страница 12: ...chien Tel 00420 47 5620808 Fax 00420 47 5620299 DSD Dental Service Dienst Jaap Wim Hr John Wolters de Jong Bosma Ketting Markerkant 13 03 i NL 1314 AL Amere Netherlands Tel 0031 36 53 4 32 12 Fax 0031...

Страница 13: ...oup 1 The Ortholux Luminous Curing Light unit uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic eq...

Страница 14: ...e 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on...

Страница 15: ...determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol N...

Страница 16: ...eparation distance according to frequency of transmitter m Rated maximum output of transmitter W 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 04 0 12 0 40 0 1 0 11 0 38 1 26 1 0 35 1...

Отзывы: