Guidance and manufacturer´s declaration – electromagnetic immunity 

The Ortholux™ Luminous Curing Light unit is intended for use in the electromagnetic environment specified 
below. The customer or the user of the Ortholux Luminous Curing Light unit should assure that it is used in 
such an environment. 

Im m u n ity te st 

IE C  6 0 6 0 1  

te st le ve l 

C o m p lia n ce  le ve l 

E le ctro m a g n e tic e n viro n m e n t - 

gu id a n ce 

Electrostatic 
discharge (ESD) 
 
IEC 61000-4-2 

± 6 kV contact 

± 8 kV air 

± 6 kV contact 

± 8 kV air 

Floors should be wood, concrete or 
ceramic tile. If floors are covered with 
synthetic material, the relative humidity 
should be at least 30 %. 

Electrostatic 
transient / burst 
 
IEC 61000-4-4 

± 2 kV for power 
supply lines 
 
± 1 kV for input/output 
lines 

± 2 kV for power 
supply lines 
 
± 1 kV for input/output 
lines 

Mains power quality should be that of a 
typical commercial or hospital 
environment. 

Surge 
 
IEC 61000-4-5 

± 1 kV differential 
mode 
 
± 2 kV common mode 

± 1 kV differential 
mode 
 
± 2 kV common mode  

Mains power quality should be that of a 
typical commercial or hospital 
environment. 

Voltage dips, short 
interruptions and 
voltage variations 
on power supply 
input lines 
 
IEC 61000-4-11 

< 5 % U

T

  

(>95 % dip in U

T

 ) 

for 0,5 cycle 
 
40 % U

 

(60 % dip in U

T

 ) 

for 5 cycles 
 
 70 % U

 

(30 % dip in U

T

 ) 

for 25 cycles 
 
< 5 % U

T

  

(>95 % dip in U

T

 ) 

for 5 sec  

< 5 % U

T

  

(>95 % dip in U

T

 ) 

for 0,5 cycle 
 
40 % U

 

(60 % dip in U

T

 ) 

for 5 cycles 
 
 70 % U

 

(30 % dip in U

T

 ) 

for 25 cycles 
 
< 5 % U

T

  

(>95 % dip in U

T

 ) 

for 5 sec 

Mains power quality should be that of a 
typical commercial or hospital 
environment. If the user of the Ortholux 
Luminous Curing Light unit requires 
continued operation during power mains 
interruptions, it is recommended that 
the Ortholux Luminous Curing Light unit 
be powered from an uninterruptible 
power supply or a battery. 

Power frequency 
(50/60 Hz) 
magnetic field 
 
IEC 61000-4-8 

3 A/m 

3 A/m 

Power frequency magnetic fields should 
be at levels characteristic of a typical 
location in a typical commercial or 
hospital environment. 

NOTE         U

T  

is the a. c. mains voltage prior to application of the test level. 

 

 

 

Содержание Ortholux Luminous Curing Ligh

Страница 1: ...ENGLISH Fast Curing Cordless LED Light Instructions For Use Ortholux Luminous Curing Light...

Страница 2: ...age stated on the rating plate is compatible with the available main voltage Operation of the unit at a different voltage may damage the unit 3 Position the unit so that the power plug is accessible a...

Страница 3: ...devices 18 Do not use Ortholux Luminous Curing Light for patients with a history of photobiological reactions including individuals with urticaria solaris or erythropoetic protoporphyria or who are c...

Страница 4: ...The handpiece switches to sleep mode once it is placed in the charger or if left unused for approximately 5 minutes outside the charger NOTE Do not discard these Instructions for the duration of produ...

Страница 5: ...o technical modification without prior notice Installation of the Unit Factory Settings The factory settings of the unit are as follows When first turned on the unit is in Extended Mode i e only the P...

Страница 6: ...ge the new battery completely for the first time The status LED showing the operating status of the charger flashes green during the charging operation see below Operating Status Display of the Charge...

Страница 7: ...om the Extended Mode Caution High intensity light always involves the production of heat When utilizing the light in Extended Mode 15s do not keep the light in one position Longer exposure in the regi...

Страница 8: ...varying arrangement of the components of the units NOTE The light intensity is independent of the battery charge level Caution Clean the testing area with a damp cloth Without applying pressure place...

Страница 9: ...y after the acoustical signal has gone off 2 minute cooling off time in the air typically corresponds to 1 minute operating time The light emission does not start when the START button is pressed an e...

Страница 10: ...defective batteries and units in accordance with local legal regulations Maintenance and Care The Ortholux Luminous Curing Light is maintenance free with the exception of cleaning the light guide and...

Страница 11: ...A PARTICULAR PURPOSE User is responsible for determining the suitability of the product for user s application If this product is defective within the warranty period your exclusive remedy and 3M Unit...

Страница 12: ...chien Tel 00420 47 5620808 Fax 00420 47 5620299 DSD Dental Service Dienst Jaap Wim Hr John Wolters de Jong Bosma Ketting Markerkant 13 03 i NL 1314 AL Amere Netherlands Tel 0031 36 53 4 32 12 Fax 0031...

Страница 13: ...oup 1 The Ortholux Luminous Curing Light unit uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic eq...

Страница 14: ...e 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on...

Страница 15: ...determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol N...

Страница 16: ...eparation distance according to frequency of transmitter m Rated maximum output of transmitter W 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 04 0 12 0 40 0 1 0 11 0 38 1 26 1 0 35 1...

Отзывы: