background image

 

English  10

ENGLISH

sterilization, the spots of dried liquid should be 

wiped off the ends of the device.

•  Adhering polymerized material should be removed 

with alcohol, a plastic spatula may help in removing 

the material.

-  Do not use any sharp or pointed tools for 

cleaning in order to protect the surface of the 

unit from scratching.

•  The protective glass of the handpiece can be 

cleaned with a soft and fluff-free cloth.

Cleaning the Charger, Handpiece, and 

 

Eye Shield

•  To disinfect all components, spray the disinfectant 

on a towel and use it to disinfect the unit. 

Do 

not spray the disinfectant directly on the 

handpiece or the charger.

-  Disinfection agents must not enter the unit!

•  Dry residual disinfectants on the charger, the 

handpiece and the eye shield with a soft and fluff-

free cloth, as they damage the plastic components.

•  Clean the charger, the handpiece and the eye 

shield with a soft cloth and, if required, a mild 

cleaning agent.

-  Solvents or abrasive cleaners may not be used in 

any case, as they damage the plastic components!

-  Cleaning agents must not enter the unit!
-  Make sure that charge contact pins remain dry 

and are not contacted by metallic or greasy 

parts. Do not bend the charge contact pins 

during drying. Wet charge contact pins will 

cause an operating error (malfunction message: 

the LED showing the operating status of the 

charger flashes red).

Storage of the Handpiece during  

Extended Periods of Non-Use

•  If the handpiece is not to be used for an extended 

period of time (e.g., during vacation) fully 

charge the battery prior to departure or keep the 

handpiece inserted in the operational charger.  

A safety switch within the battery prevents a  

total discharge.

•  Discharged or nearly discharged batteries must be 

recharged as soon as possible.

Disposal

Your new unit contains a powerful lithium-ion 

rechargeable battery. Dispose of defective batteries  

and units in accordance with local legal regulations!

Maintenance and Care

The Ortholux Luminous Curing Light is maintenance-

free with the exception of cleaning the light guide and 

replacing batteries as needed. The light output should 

be checked on a regular basis, i.e. daily, using the 

integrated light intensity testing area on the charging 

base. See the information contained in this chapter 

to secure problem-free operation. There are no user 

serviceable parts on the unit. If the unit malfunctions 

and the trouble shooting steps above do not remedy the 

situation, call 3M Unitek Customer Service.

Inserting/Removing the Battery 

Never place the handpiece in the charger without the 

battery inserted in the handpiece!
Use 3M Unitek batteries only! The use of other 

manufacturer batteries or non-rechargeable batteries/

primary batteries is a potential hazard and may 

damage the unit.
•  Remove the battery from the handpiece by turning 

it counter-clockwise as shown. (Page 6)

•  Insert the (new) battery into the handpiece and turn 

it clockwise until the seal is firmly pressed against 

the metal housing.

•  Place the handpiece in the charger for 1.5 hours in 

order to fully charge the new battery for the first time.

-  The LED showing the operating status on the 

charger flashes green; see also “Operating 

Status Display of the Charger.”

Handpiece/Battery Care

•  Use only the 3M Unitek charger which is provided 

with the unit; failure to do so could lead to damage 

to the battery!

•  Do not immerse the battery in water or incinerate! 

Please also observe the chapter on ”Safety.”

Cleaning the Light Guide

The light guide can be autoclaved. 

Do not sterilize by chemical means or hot-air!

•  The light guide should be regularly wiped clean 

with a soft cloth. Especially before and after steam 

Error

Cause

•  Solution

The LED showing the 

operating status in the 

charger glows red steadily.

The charge contact pins in 

the handpiece or the charger 

are wet.
•  Dry the charge contact 

pins. Be careful not to 

bend the flexible  

charge contact pins in 

the charger.

Содержание Ortholux Luminous Curing Ligh

Страница 1: ...ENGLISH Fast Curing Cordless LED Light Instructions For Use Ortholux Luminous Curing Light...

Страница 2: ...age stated on the rating plate is compatible with the available main voltage Operation of the unit at a different voltage may damage the unit 3 Position the unit so that the power plug is accessible a...

Страница 3: ...devices 18 Do not use Ortholux Luminous Curing Light for patients with a history of photobiological reactions including individuals with urticaria solaris or erythropoetic protoporphyria or who are c...

Страница 4: ...The handpiece switches to sleep mode once it is placed in the charger or if left unused for approximately 5 minutes outside the charger NOTE Do not discard these Instructions for the duration of produ...

Страница 5: ...o technical modification without prior notice Installation of the Unit Factory Settings The factory settings of the unit are as follows When first turned on the unit is in Extended Mode i e only the P...

Страница 6: ...ge the new battery completely for the first time The status LED showing the operating status of the charger flashes green during the charging operation see below Operating Status Display of the Charge...

Страница 7: ...om the Extended Mode Caution High intensity light always involves the production of heat When utilizing the light in Extended Mode 15s do not keep the light in one position Longer exposure in the regi...

Страница 8: ...varying arrangement of the components of the units NOTE The light intensity is independent of the battery charge level Caution Clean the testing area with a damp cloth Without applying pressure place...

Страница 9: ...y after the acoustical signal has gone off 2 minute cooling off time in the air typically corresponds to 1 minute operating time The light emission does not start when the START button is pressed an e...

Страница 10: ...defective batteries and units in accordance with local legal regulations Maintenance and Care The Ortholux Luminous Curing Light is maintenance free with the exception of cleaning the light guide and...

Страница 11: ...A PARTICULAR PURPOSE User is responsible for determining the suitability of the product for user s application If this product is defective within the warranty period your exclusive remedy and 3M Unit...

Страница 12: ...chien Tel 00420 47 5620808 Fax 00420 47 5620299 DSD Dental Service Dienst Jaap Wim Hr John Wolters de Jong Bosma Ketting Markerkant 13 03 i NL 1314 AL Amere Netherlands Tel 0031 36 53 4 32 12 Fax 0031...

Страница 13: ...oup 1 The Ortholux Luminous Curing Light unit uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic eq...

Страница 14: ...e 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on...

Страница 15: ...determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol N...

Страница 16: ...eparation distance according to frequency of transmitter m Rated maximum output of transmitter W 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 04 0 12 0 40 0 1 0 11 0 38 1 26 1 0 35 1...