![3Disc Heron IOS Скачать руководство пользователя страница 36](http://html.mh-extra.com/html/3disc/heron-ios/heron-ios_user-manual_442192036.webp)
Page 36
User Manual
5.4
Recommended Scan Workflow
Before launching a scan,
make sure the
Heron
TM
IOS Scanner is correctly connected to
the computer via the USB 3.0 cable and the HeronClinic™ software is running. Make
sure the computer on which the HeronClinicTM software is installed is connected to
an external power source.
See above:
Getting Started With HeronClinic
Useful Scanning Tips
Always take care when scanning to:
•
Stay in contact with or close to the teeth
(recommended distance from tip to tooth is 0-12 mm)
•
Scan slowly and smoothly, to avoid interrupting the 3D reconstruction
(minimum 1 second per tooth)
•
Keep teeth in the center of the field of view
•
Avoid scanning lips, cheeks, tongues, gloves, etc.
•
Do not hesitate to reset the scan if necessary
Begin your scan by starting on the arch of chosen restoration. If you have selected to
scan the antagonist, you may start by scanning either arch.
5.4.1
Scanning the Maxillary Arch
To make the scanning process
more
comfortable
for maxillary scanning, you
may remove the tip by pressing the button
(1) located on the bottom of the handpiece
and rotating the scanner tip 180 degrees
(2).
IMPORTANT
: make sure the computer on which the HeronClinic
TM
software is installed is connected to an external power source
before starting to scan with the Heron
TM
IOS.
Do not scan using a battery-powered computer.
Содержание Heron IOS
Страница 1: ...IOS FPM 71 001 3 1_EN User Manual Version 3 4 x ...
Страница 3: ......
Страница 54: ...Page 54 User Manual 4 Select the desired options 5 Click Save to save selected View Notification settings ...
Страница 68: ...Page 68 User Manual INMETRO Certification Mark Ukraine Conformity Mark ...