3Disc Heron IOS Скачать руководство пользователя страница 8

7

7. 

The scan is saved automatically at each step in the workflow.

8. 

To restart the scan, press the reset button. This will delete the current scan and 

allow one to start a new scan.

9. 

After a scan step is complete you will see a green check mark in the workflow at 

the bottom of the screen beneath the appropriate icon.

Scanning Step 4: Finalization (A.T.R.S)

After the basic scan has been conducted, below are some tips that should be followed 

to finalize the scan.

 

Analyze

 the scan for completeness and accuracy, make sure to rotate the scan 

image on the main view screen in order to obtain a complete view of the scan 

data

 

Trim

 away excess information (tongue, cheek, vestibule, tissue, fingers, etc.) 

that may interfere with a successful bite alignment. (The Trim options are on 

the right side work-flow in the “TOOLS” icon)

 

Rescan

 any area that is missing critical data for the success of the final 

restoration

 

Save

 and proceed to the next step in the scan strategy

Scanning Step 5: Exporting Files with Dentalshare & STL

• 

After you have successfully scanned the upper and lower jaws and aligned the 

bite it is time to export the scan files to the dental laboratory for fabrication of 

the restoration

• 

Dentalshare is a built-in file sharing system, your dental laboratory needs to 

sign up for a dentalshare account in order to use the dentalshare network

• 

STL exporting (outside of dentalshare)

Keyboard Shortcuts

Use these keys anytime:

• {R} – reset entire scan

• {X} – re-center scan

• {Space} – start & stop scanning

• {Enter} – same as clicking OK

• {Escape} – same as clicking CANCEL

To use while navigating:

• {Ctrl + Z} – undo last step

• {Ctrl + Y} – redo last undo

• 

{+/=} – increase size of tool

• 

{-/_} – decrease size of tool

• {C/Insert} – Crop to current 

selection

• {D/Delete} – Delete selection

• 

Hold Ctrl while clicking a tooth to 

copy last selection

• 

Hold Shift to apply last setting to 

multiple teeth

Содержание Heron IOS

Страница 1: ...g the Tip 2 Patient Environment 2 Tip Cleaning Sterilization 3 Scanning Step 1 Creating a Case 5 Scanning Step 2 Scanning Preparation 5 Scanning Step 3 Basic Scanning Procedure 6 Scanning Step 4 Final...

Страница 2: ...ng station on a flat stable surface and place the Heron IOS hand piece securely in its docking station Step 2 Connect the AC DC power adapter cable to the docking station the connector socket is locat...

Страница 3: ...ine extending 1 8 meter 6 feet around patient and system Depress bottom button 1 and rotate tip at 90 increments 2 Tip should be locked into position prior to scanning 1 2 1 2 Attach the tip to the ha...

Страница 4: ...following procedure Step 1 Clean the tip with soap water ensuring that the mirror is clean and free of smudges stains or any milky haze Do not use an abrasive surface to clean the mirror Step 2 After...

Страница 5: ...Insert and seal the tip into a sterilization pouch Step 2 Sterilize the wrapped tip in a steam autoclave at a temperature of 132 C 270 F and 210kPa 30 46 psi for 6 minutes 132 C 270 F 210kPa 30 46psi...

Страница 6: ...Click the yellow SCAN icon in top right corner of screen 7 A B E Quantor3D will auto launch taking you to the scan window Scanning Step 2 Scanning Preparation Prepare the patient for the scan by ensur...

Страница 7: ...ns are obtained at a closer range and it is okay if the tip touches the teeth During scanning you should continue to hear a rapid clicking sound and observe a green rectangle indicating the last scann...

Страница 8: ...Rescan any area that is missing critical data for the success of the final restoration Save and proceed to the next step in the scan strategy Scanning Step 5 Exporting Files with Dentalshare STL After...

Страница 9: ...can scanning is a skill that requires practice Prior to use on a patient practice should be conducted using a dental typodont model Lower Upper Jaw Scans 1 Start the scan path on the working arch jaw...

Страница 10: ...g the buccal surface of the posterior teeth and the facial surface of the anterior teeth to complete your scan strategy 5 You can turn the scanner off at any point during the scan make sure you restar...

Страница 11: ...will ring and a green check mark will appear on the upper jaw image in the upper right corner of the application 2 Once the upper jaw is aligned move the handpiece down to the lower jaw Again move th...

Страница 12: ...d do not throw it away The box is optimal for any necessary transportation or shipping of the Heron IOS 3D Imaging and Simulation Corp Americas 22560 Glenn Drive Suite 116 Sterling VA 20164 USA Phone...

Отзывы: