33
CG000396 | Rev C
User Guide | Chromium X Series (X/iX) | Troubleshooting
The touchscreen will guide the user through recoverable errors. If the
error continues, or if the instrument has seen critical or intermediate
errors, email [email protected] with the error code displayed.
Support will request log files for troubleshooting. Upload pertinent logs
to 10x Genomics by navigating to the Logs menu option on screen. (See
There are two types of errors:
Critical Errors
– When the instrument has seen a critical error, the run
will immediately abort. Do not proceed with any further runs. Contact
[email protected] with the error code.
• System Error
• Pressure Error
• Chip Error
• Run Error
• Temperature Error
• Firmware Error
User Recoverable Errors
– Follow error handling instructions through
the touchscreen and continue the run.
• Gasket Error
• Tray Error
• Chip Error
• Unsupported Chip Error
• Network Error
• Update Error
This factory button aids qualified 10x Genomics service personnel in
manually ejecting the instrument tray in the event that the instrument is
unresponsive.
DO NOT open the Chromium X/iX. There are no user-serviceable parts
inside. Refer all servicing to qualified 10x Genomics service personnel.
Error Screens
Pressure Release
Valve
Содержание Chromium X Series
Страница 5: ...Introduction 6 Introduction 7 Product Identification 7 Product Specifications ...
Страница 8: ...Safety Compliance Information 9 Safety 10 Regulatory ...
Страница 11: ...System Components 12 System Components 13 Chromium X iX Accessory Kits ...
Страница 14: ...System Operation 15 Installation 17 First Time Use 18 Readiness Test 19 Main One Touch Operation ...
Страница 21: ...Tips Best Practices 22 Chromium Next GEM Chip 24 Chip Placement on Chromium X Tray 24 10x Genomics Cloud ...
Страница 25: ...Menu Options 26 Menu Options 28 Firmware Update Procedure ...
Страница 29: ...Maintenance 30 System Maintenance 30 Moving Shipping the Instrument ...
Страница 32: ...Troubleshooting 33 Error Screens ...