INSTRUCTION MANUAL

Volt CPAP Battery Pack

PLEASE READ THIS SHEET CAREFULLY TO ENSURE SAFE 

USAGE AND MAINTENANCE

The Viverity

 Volt CPAP Battery Pack 

provides you with a simple solution at just 
over 2 lbs, including carrying case. It comes 
complete with a large capacity battery, 
AC charger and handy carrying case. The 
carrying case has a handle, shoulder strap 
and four external zipper storage pouches to 
give you a wide range of storage capabilities. 
This unit will provide power to your CPAP 
unit for up to 14 hours

*

 of operation.

NOTE

The Volt CPAP Battery Pack requires  

a manufacturer’s DC auto adapter for  

your individual CPAP/Bi-PAP unit 

(purchased separately). See your  

Homecare Provider or the original 

manufacturer of your CPAP unit.

Use:

1.   Remove the charger from the case  

and plug the male 12 volt end of the 
charger into the female 12 volt outlet  
on the CPAP battery.

2.   Charge the battery until the light on the 

charger turns green or a minimum of 6 
hours on the first charge. A red light on 
the charger indicates that the battery is 
still charging.

3.   When fully charged, use the 

manufacturer’s DC cord (not included)  
to connect your CPAP to the outlet on 
the Volt CPAP Battery Pack. 

NOTE

Most heated humidifiers will disable while 

on battery operation due to excessive 

battery consumption. The humidifier can 

be used in cold passover mode instead 

while battery is in use.

Always:

•   Always charge the battery fully prior to 

use. Only use the provided charger for 
charging your Volt CPAP Battery Pack.

•   Always use your Volt CPAP Battery Pack 

and charger in well ventilated areas.

•    Always allow the battery to cool to room 

temperature prior to charging, in case 
it has been heated while stored in a car, 
truck, etc.

•   Always disconnect the CPAP unit from 

the battery when not in use. The displays 
on the units consume power, shortening 
your usage time.

•   Always remove the Volt CPAP Battery 

Pack from the case when charging. It is 
normal for it to feel warm when charging 
a very low battery. 

•   Always stop using the battery if a leak 

is detected or if the unit becomes 
extremely hot.

Never:

•   Never try to open or pierce the wrap 

around the batteries to service it.

•   Never try to service the battery in any 

way. It is factory sealed and changes may 
cause damage to the unit or your CPAP.

•   Never replace the batteries individually or 

use in conjunction with another charger, 
as damage can occur.

•   Never submerge/allow batteries to 

become wet. 

•   Never subject the Volt CPAP Battery Pack 

to temperatures over 140° F.

•   Never allow children to play with your 

Volt CPAP Battery Pack, as it poses a 
choking and electrical hazard.

•   Never throw away or incinerate your 

Volt CPAP Battery Pack. Please dispose 
of the battery according to local waste 
guidelines for recycling.

•   Never place the battery in a microwave  

or pressurized container.

•    Never place the Volt CPAP Battery Pack 

in checked luggage of any aircraft. It is 
safe to include with carry-on bags.

Battery Type

Li-Ion 18650

Battery Pack

14.8 VAC

Charger

15.6 VDC-3A OUTPUT, 100-240 VAC INPUT, with 
Charge Indicator and Male 12 Volt Auto Connector

Weight

2.4 lbs.

Rated Capacity

7,800 mAh

Minimum Capacity

7,500 mAh

Rated Energy

115.44 Watts

Loaded Voltage

14.4 Volts

Cut Off Voltage

11.00 Volts

Charge Voltage

15.6 Volts

Maximum Charge Current

3,900 mA

Maximum Discharge Current

1,560 mA

Operating Temperature

32˚F – 113˚F (0˚C – 45˚C)

Storage Temperature

32˚F – 113˚F (0˚C – 45˚C)

Roscoe Medical, Inc.

Strongsville, OH 44149
800.376.7263 • 

roscoemedical.com

A division of Compass Health Brands
Made in China ©2015

Operation

Hours

*

cm H

2

0

(up to) 9

15 

(up to) 14

10