background image

   

PRINCIPLE AND INTENDED USE 

 
The  VivaChek™  Lancing  Device  is  used  with  VivaChek™  sterile  lancets  or  any  other  compatible 
disposable sterile lancets to draw capillary blood from the fingertip, palm (at base of the thumb) or 
forearm, for blood glucose testing or other testing utilizing small amounts of blood. 

INSTRUCTIONS FOR USE 

Before testing, choose a clean, dry work surface. Familiarize yourself with the procedure and make 
sure you have all the items needed to obtain a drop of blood. 

 

Important: Prior to testing, wipe the test site with an alcohol swab or soapy water. Use warm water to 

increase blood flow if necessary. Then dry your hands and the test site thoroughly. Make sure there is 

no alcohol, soap or lotion on the test site. 

  SAMPLING FROM THE FINGERTIP 

For fingertip sampling, adjust the depth penetration to reduce the discomfort. You do not need the 
clear cap for fingertip sampling. 
1.  Unscrew the lancing device cover from the 

body of the lancing device. Insert a sterile 
lancet into the lancing  device and push it 
until the lancet comes to a complete stop 
in the lancing device. 

2.  Hold the lancet firmly in the lancing device 

and twist the safety tab of the lancet until it 
loosens,  then  pull  the  safety  tab  off  the 
lancet.  Save  the  safety  tab  for  lancet 
disposal.   

3.  Carefully  screw  the  cover  back  onto  the 

lancing  device.  Avoid  contact  with  the 
exposed  needle.  Make  sure  the  cover  is 
fully sealed on the lancing device. 

4.  Adjust  the  puncture  depth  by  rotating  the 

lancing device cover. There are a total of 5 
puncture depth settings. To reduce  the discomfort, use the  lowest setting  that  still  produces  an 
adequate drop of blood. 

Adjustment:   

1    for delicate skin   
2 and 3    for normal skin   
4 and 5    for calloused or thick skin 

Note: Greater pressure of the lancing device against the finger will also increase the puncture depth.

 

 

 

5.  Pull the cocking barrel back to set the lancing device. You may hear a 

click to indicate the lancing device is now loaded and ready for obtaining 
a drop of blood.   

6.  Prior  to  testing,  wipe  your  hand  with  an  alcohol  swab  or  wash  your 

hands with soap. Use warm water to increase 
blood  flow  in  your  fingers  if  necessary.  Then 
dry your hands thoroughly. Massage the hand 
from the wrist up to the fingertip a few times to 
encourage blood flow.       

7.  Hold the lancing device against the side of the 

finger  to  be  lanced  with  the  cover  resting  on  the  finger.  Push  the  release  button  to  prick  your 
fingertip. You should hear a click as the lancing 
device  activates.  Gently  massage  from  the 
base  of  the  finger  to  the  tip  of  the  finger  to 
obtain  the  required  blood  volume.  Avoid 
smearing the drop of blood.     

Note: For the greatest reduction in pain, lance on the sides of the fingertips. Rotation of sites is 

recommended. Repeated punctures in the same spot can make your fingers sore and callused. 

8.  Unscrew  the  lancing  device  cover.  Place  the  safety  tab  of  the  lancet  on  a  hard  surface  and 

carefully insert the lancet needle into the safety tab. 

9.  Press  the  release  button  to  make  sure 

that the lancet is in the extended position. 
Slide  the  ejection  button  forward  to 
discard the used lancet to a container for 
sharp  objects.  Place  the  lancing  device 
cover back on the lancing device. 

Warning:  

The VivaChek™ lancing device is designed for self testing by a single 

patient. It must not be used on more than one person due to the risk of 

infection.  

Use a new VivaChek™ lancet each time you test because it is no 

longer sterile after use.   

Always dispose of the used lance properly to prevent injury or contamination to others. 

SAMPLING FROM FOREARM OR PALM (AT THE BASE OF THE THUMB) 

 
The forearm and palm areas have less nerve endings than the fingertip. You may find that obtaining 
blood  from  these  sites  is  less  painful  than  from  the  fingertip.  The  procedure  for  forearm  and  palm 
sampling is different.  You  need the clear cap to draw blood from these sites.  The clear cap  is not 
adjustable for puncture depth.   
Follow Step 1 and 2 to insert the lancet and load the lancing device. 
 
3.  Screw  the  clear  cap  onto  the  lancing 

device. 

Note: To bring fresh blood to the surface of 

the puncture site, massage the puncture site 

vigorously for a few seconds until you feel it 

getting warm 

4.  Choose a puncture site on the forearm or palm. Select a soft and fleshy area of the forearm and 

palm that is clean and dry, away from bone, and free of visible veins 
and hair. 

5.  Place the lancing device against the puncture site. Press and hold 

the clear cap against the puncture site for a few seconds. Press the 
release button of the lancing device, but do not immediately lift the 
lancing device from the puncture site. Continue to hold the lancing 
device  against  the  puncture  site  until  you  can  confirm  a  sufficient 
blood sample has formed. 

 

  CARE OF THE LANCING DEVICE 

 
Clean  the  lancing  device  using  a  soft  cloth  with  mild  soap  and  warm  water  as  required.  Use  70% 
Isopropyl Alcohol to disinfect the lancing device. Carefully dry the lancing device. Do not immerse the 
lancing device in liquid. 

PRECAUTIONS 

  Do not use the lancet if the safety tab is missing or loose when you take the lancet out of the bag. 

  Do not use the lancet if the needle is bent. 

  Use caution whenever the lancet needle is exposed. 

  Never share lancets or the lancing device with other people. 

  In order to reduce the risk of infection from prior use of the instrument, always use a new, sterile 

lancet. Do not reuse lancets. 

  Avoid getting the lancing device or lancets dirty with hand lotion, oils, dirt or debris. 

LIMITED WARRANTY 

If the lancing device does not work for any reason other than obvious abuse in the first 5 years after 
purchase, we will replace it with a new or equivalent lancing device free of charge. Please contact 
your local dealer for replacement. 

INDEX OF SYMBOLS 

 

 

Consult instructions for 
use 

 

Lot number 

REF  

Tuotenume 
Catalog number 

 

Manufacturer

 

 

Authorized Representative 

  

 

 

VivaChek™ is a trademark of VivaChek Biotech (Hangzhou) Co., Ltd.  Number:

 

1201021303 

© 2021 VivaChek Biotech (Hangzhou) Co., Ltd. 

Effective date: 2021-07-23

                                                                                                                                                                             

MODEL HH-XIII-T    REF VGD01-001     

English 

 

Lancing device package insert

Reviews: