WARNING - READ BEFORE INSTALLATION 

Vaisala Differential Pressure Transmitter PDT102 is a high-performance 

instrument designed primarily for use in life science and cleanroom 

applications. PDT102 is not authorized for use as a critical component in life 

support devices or systems. Consult Vaisala before installing if there are any 

questions or concerns. 

Overpressure

: Pressure spikes in excess of the rated overpressure capability 

of the transmitter may cause irreversible electrical and/or mechanical damage 

to the transmitter. 

Static electrical charges

: To avoid damage to the transmitter the 

operator/installer should follow proper ESD (electrostatic discharge) 

protection procedures. 

TECHNICAL SPECIFICATION 

Performance 

 

Measurement range (bidirectional) 

±50 Pa or  

±0.25 in H

2

Overpressure 

Proof pressure 

0.7 bar 

 

Burst pressure 

1.7 bar 

 

Static pressure 

1.7 bar 

Pressure type 

Differential, gauge, vacuum and compound 

Accuracy 

(incl. non-linearity, hysteresis, 

repeatability and zero/span 

calibration settings) 

0.25 % span 

or 0.5 % span, 

depending on choice 

Repeatability 

for 0.25 % span accuracy 

0.03 % 

 

for 0.5 % span accuracy 

0.05 % 

Electrical resolution 

 

1 x 10

-4

 span 

Long-term stability 

 

≤0.5 % span/year 

Response time  (10...90 %) 

250 ms 

Warm-up time 

 

15 s 

Compensated temperature range 

 

+2...+57 °C 

(+35.6...+134.6 °F) 

Temperature dependence 

±(0.036 Pa + 0.036 % of reading)/°C 

or  

±(0.0001 inH

2

O +  0.02% of reading)/°F   

(reference 21 °C or 70 °F) 

Mounting position error (zero adjustable) 

≤0.25 % 

Adjustments (front accessible) 

Zero (span low) 

±5 % span 

 

Span (span high) 

±3 % span 

Operating environment 

Operating temperature 

-29...+70 °C (-20.2...+158 °F) 

Storage temperature 

-40...+82 °C (-40...+179.6 °F) 

Electromagnetic compatibility 

EN61326-1, Basic immunity  

test requirements 

Note: If used in an electromagnetic field of 3 V/m, with narrow frequency area 

of 80 - 120 Mhz, it is possible that the current output of PDT102 can deviate  

max. 0.3 % (with accuracy specified 0.25%). 

Inputs and outputs 

Process connection 

1/8 NPT female according to ANSI/ASME B1.20.1 

Output signal 

2-wire 

4...20 mA 

 

3-wire 

0...5 V 

Operating voltage 

 

12...36 VDC 

Max. loop resistance for 4...20 mA 

≤ (Supply voltage - 12V)/0.022 A 

Supply current 

for 0...5 V output 

max. 10 mA 

 

for 4...20 mA output 

max. 20 mA 

Electrical connection 

Screw terminals, 12...22 AWG (0.33 up to 3.31 mm

2

Mechanics 

Medium (measured gas) 

Clean and dry air, non-conducting and 

non-corrosive gases 

Material 

Process connection 

Brass 

 

Sensor element 

Silicon, aluminum, glass 

 

Case 

Polycarbonate, glass filled (UL94-V-1) 

Mounting 

DIN rail EN 50022, EN 50035  and EN 50045 

Housing classification 

IP30 

Weight 

 

0.16 kg 

MOUNTING 

When securing to EN50022 rail first hook the top portion of the rear clip onto 

the top of the rail and push gently into place. To remove, simply place finger 

behind lower rear lever and pull forward. To remove from the EN50035 rail, 

grasp the unit, push up, and rotate the bottom out. For the EN50045, pull 

down and rotate the bottom out. 
Multiple units can be stacked across the rail. Internal circuitry is vented 

through the bottom of each unit. Tag holes are fabricated into the lower rear 

lever and the housing below the terminal strip for calibration reference tagging 

and/or ID location. 

125.99

Span

Zero

V+    V–    Out

22.6

99.6

80.8

 

Figure 1

 

Dimensions (in mm) 

134.4

Span

Zero

V+    V–    Out

22.6

99.6

80.8

Test:

+

-

 

Figure 2

 

Dimensions with Process Valve Actuator and 

Test Jacks (in mm) 