background image

 

 

 

 

 

 

 

Digital Strain Indicator

 

 

TD−35

 

Operation Manual

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Teac Corporation 

                                   

     

   

 Ver

 

 2006/ 4/18 

              

                             

  

 

P/N

 

D009185‑00C

1

 

Summary of Contents for TD-35

Page 1: ...Digital Strain Indicator Operation Manual Teac Corporation Ver 2006 4 18 P N D009185 00C 1...

Page 2: ...e of this product according to the service standard prescribed by TEAC Corporation This TEAC warrantee shall also fully comply with any local product liability law Teac Corporation Headquarter Sales D...

Page 3: ...Index 1 Overview 4 2 Features 4 3 Configuration 4 4 Standard Accessories 6 5 Specifications 6 6 Operation 8 7 Measuring Method 12 8 Miscellaneous Materials 15 9 External View 18 3...

Page 4: ...ut value mV V for direct reading convenience This operates with a built in battery AAA type or external AC power supply using an AC adaptor 2 Features 1 Small and compact size 2 All necessary function...

Page 5: ...5...

Page 6: ...lay module LCD Character height approx 12mm Display frequency Approx 6 times second Peak hold Maximum value during measurement Calibration accuracy 0 1 F S or less FS 20000 10 strain Linearity 1 10 st...

Page 7: ...Dimension 150mm W x100mm D x40mm H excluding umbo Weight Approx 500g including ZR6 Y dry cell battery 7...

Page 8: ...for the display back light Note that twice as much power is consumed when it is turned on in battery operation INDICATION Display Unit Switch The switch to select whether the strain input is displaye...

Page 9: ...the strain gauge or strain gauge sensor INPUT Sensor Input Connector The NDIS standard input connector for the strain gauge sensor Applicable plug model is PRC03 12A10 7M10 5 made by Tajimi Musen Pin...

Page 10: ...ey Key Used to move the digit while setting value ENTER Key Used to set the value in various settings PEAK HOLD Key Used to select the peak hold operation It holds and displays the maximum value of th...

Page 11: ...in the specification is under the conditions of a ZR6 Y Oxiride dry cell battery used at room temperature When the voltage drops below 4V B is displayed on the LCD When it occurs replace the battery...

Page 12: ...n may occur such as rapid temperature or humidity changes 7 2 Wire Connection This is a strain indicator for measuring the strain gauge type sensor gauge output such as a load cell For the strain meas...

Page 13: ...he rated capacity is a value found in the sensor s data sheet To change the decimal point position move the cursor to a desired decimal point position using or key press ENTER key and the character di...

Page 14: ...range Countermeasure Refer to Zero Adjustment in the back of the material and adjust the bridge balance to the level smaller than 10000 10 strain AD OVER OVER Description The input level from sensor e...

Page 15: ...weight is generated The zero adjustment range of TD 35 is 5mV V 10000 10 strain In case it does not meet the zero adjustment within this range as shown in the figure below the zero point shift resiste...

Page 16: ...connect an appropriate resister RZ to one side of the bridge and record the displayed value then remove the resister When the indicator is replaced for a reason of servicing or so forth connect the sa...

Page 17: ...Added resister value versus input conversion strain of the bridge circuit with 350 and 120 resistance List of switch settings on measurement items 17...

Page 18: ...9 External View MADE IN JAPAN NO SERIAL MODEL NO TEAC INSTRUMENTS CORP DIGITAL STRAIN INDICATOR 5 12V 0 3 0 14A 18...