background image

 

 

2842 Business Park Ave. 
Fresno, California 93727-1328 
(559) 292-2171

 (800) 456-8168 

112675 REV.A 

 

April 2009 

 

Page 1 of 2 

INSTRUCTION SHEET 

Pulse Rnet Controller Replacement Instructions

 

FOR DEALER USE  Installed by: 

Date: 

(retain as record of installation) 

 
 
 
 

Note: All programming of profiles and preferences are stored within the motor controller.  It is recommended that the program 

be “Read” or save prior to 

replacement.  Once the controller has been replaced the stored program can be “Written” or loaded back into the controller.  This process can be accomplished 
with a DTT or PC programmer.  
 

1.  Disconnect battery power prior to servicing the motor controller.    

a.  Carefully remove seat. (See owners manual Sec. I. Seat Removal). 
b.  Remove the shroud by first removing the six fasteners (Fig. 1) using a flat head screwdriver as shown (Fig. 2). Then lift up on the shroud. 
c.  Unplug batteries from the chair (Fig. 3).  

 

2.  The controller is mounted at the rear of the chair (Fig. 4).  Firmly lift up on the controller.  The controller mounts securely using 2 locating pins.  
 
3.  Unplug the 2 motor connectors, battery connector, bus connector, and inhibit/actuator connector. (Connectors have locking latches that require releasing 

by squeezing with fingers while pulling connectors.) 

 

  

 

  

  

   

                              

Fig. 1 

 

 

Fig. 2 

 

 

 

Fig. 3 

 

 

 

Fig. 4 

4.  Remove the controller.  

 

5.  Your replacement controller will arrive with the mounting plate already mounted.  Install replacement controller by reversing instructions.  (Ensure that 

connectors are firmly inserted until the latch “clicks” into place.  

 

6. 

Once the controller is fully installed power the controller on via the remote or OMNI.  (There will be an initialization “hour-glass” then the chair will be 
ready to use).   

 

Connectors 