Sterling Sterilisers

Acuatic Vacuum 12 B

User Instructions

Summary of Contents for Acuatic Vacuum 12 B

Page 1: ...Sterling Sterilisers Acuatic Vacuum 12 B User Instructions...

Page 2: ...n EC No 1882 2003 0086 indicates that the Notified Body employed to check conformation with the above is BSI This symbol indicates that the number next to it is the serial number of the steriliser Thi...

Page 3: ...Logger Daily Maintenance 4 DailyTesting 4 Weekly Testing 4 Servicing 4 Annual Inspection 4 Annual Testing 4 Getting Started GettingStarted 5 Filling 5 Water Quality 5 Draining 5 Options 5 Operation S...

Page 4: ...onstruction structural parts are manufactured from stainless steel or aluminium Impossible to overheat innovative engineering eliminates downtime due to overheating Easy water fill just pour into the...

Page 5: ...ising Drying 138 C 134 C 1 0 2 Pressure bar 1 Time Air removal and heating Drying Sterilising B type A door seal B chamber with tray rack and trays C water filling spout D touch screen display E clean...

Page 6: ...d instruments the internal surfaces of the steriliser and the steriliser door may be hot Do not touch these parts until cooling has taken place and exercise care when handling just sterilised instrume...

Page 7: ...is observed Weekly Testing Run the automatic air leakage test This test will show that the steriliser is capable of pulling and holding a certain level of vacuum The display will indicate a pass or f...

Page 8: ...this is not recommended because of the risk of the heaters becoming scaled Draining When the used water reservoir is full the display will ask for it to be drained This will need to be done before the...

Page 9: ...If the N type cycle is enabled you will be offered a choice of starting N type or B type cycles Processing When a cycle is running the display will show the cycle time elapsed the operating temperatu...

Page 10: ...ep the Add shortcut to desktop checkbox ticked To import cycle data from the SD memory card to your computer run the Securilog application by double clicking the Securilog icon on your desktop Remove...

Page 11: ...with the Medical Devices Directive 4 A Product Registration and Customer Survey form Please take the time to print this out complete it and return it to Sterling Sterilisers so that we can keep our re...

Page 12: ...apped pouched or not Limits Design pressure 1 to 3 bar Design temperature 2 to 144 C Water Water used per cycle 200 to 700 ml Clean water reservoir volume 2 litres Used water reservoir volume 2 litres...