background image

Bluetooth

® 

Datalogging 

RH / Temperature /

Barometer

800018

Instruction Manual

Summary of Contents for 800018

Page 1: ...Bluetooth Datalogging RH Temperature Barometer 800018 Instruction Manual ...

Page 2: ...INTRODUCTION 3 2 FEATURES 4 3 MATERIALS SUPPLIED 4 4 MONITOR DESCRIPTION 5 5 SETUP 6 6 PROGRAMMING THE DEVICE 7 7 DOWNLOADING THE DATA 11 8 ADDITIONAL FEATURES 12 9 CARE AND MAINTENANCE 13 10 SPECIFICATIONS 14 11 WARRANTY 15 ...

Page 3: ...hes this warning from the phone Measurement intervals can be set up to 12 times per day with 16000 datapoints available per parameter Once the dataset is complete there is no need to attach the device to a computer to retrieve comprehensive reports Simply download a full color graphical report and e mail it directly from the phone to any computer for printing or further analysis in Excel Quickly v...

Page 4: ...4 FEATURES Min Max for each function Attached flip out tag for wall mounting Magnetic backing Long life Lithium Battery MATERIALS SUPPLIED Temperature CR2 Lithium ion battery Instruction manual ...

Page 5: ...5 MONITOR DESCRIPTION 1 LCD Display 2 Color coded LED light 3 Power Start Stop 4 Min Max 5 Magnets for mounting 6 Flip out wall mount 7 Battery Door ...

Page 6: ...larity 3 Replace the battery door and turn clock wise using gentle inward pressure Download the App You will need to download the free App from iOS or Android before using this device 1 Search Nice Logger in the App store and look for this icon 2 Download the App and open it on your smartphone 3 Connect the phone to the logger by select ing the serial number on the screen that matches the logger ...

Page 7: ...ing the Device can provide a description for all reports such as warehouse temperature or My datalogger PROGRAMMING THE DEVICE Once the device starts logging you will not be able to change the settings Make sure all parameters are set properly before starting the device 1 Press POWER to turn the device on 2 Open the app on your smart phone and connect to the device by pressing the serial number in...

Page 8: ...ect the starting mode as key or on time All methods require you to push the start button on the App before the logging will commence If the logging is for a future date you will still need to press start but the logging won t begin until the date and time you set 8 Set the stop time There are four options for this which are described below Memory Full or Key stop The logger will continue taking re...

Page 9: ...e time you set You can override this by manually stopping the logger at any time Memory Loop till Date Time The logger will continue taking readings until the date and time you set If the memory is full before the preset date and time it will continue taking readings and replace the earliest data with new readings You can override this by manually stopping the logger at any time 9 Set the alarm to...

Page 10: ...e Logger ID sec tion on the main screen standard or metric and press done 13 Press the Start Logging button on the main screen of the phone to begin The meter itself will display REC to confirm that the datalogging is underway IMPORTANT DO NOT PRESS START STOP ON THE DEVICE IT WILL SHUT THE MONITOR OFF IF YOU SHUT THE MONITOR OFF YOU WILL NEED TO RESAVE THE SETTINGS BEFORE INITIATING THE DATALOGGI...

Page 11: ...number of the device in the gray box and click Done 5 The App will download the data from the device as indicated by the progress bar 6 After the file has downloaded return to the main operation screen and click files at the bottom of the logger screen All downloaded files will appear 7 Select the file in question and choose CSV Excel or PDF format from the icons at the top of the screen 8 Your fi...

Page 12: ... screen and toggle the password option on Manual Min Max You can view the minimum and maximums for each parameter at any time directly on the device without having to look at the phone app Simply press the min max button to toggle through the minimum and maximum values Once they have been viewed toggle to the main screen by pressing min max again ...

Page 13: ...ing carbon alcohol or benzenes on the monitor Do not drop the monitor Repairs or services not covered in this manual should be performed by qualified personnel only Please contact Sper Scientific to speak with a technician Use only the recommended battery when replacement is needed ...

Page 14: ...H 0 1 to 80 0 1 3 25 C 10 to 80 RH otherwise 5 Barometric pressure 300 to 1100 hPa 8 9 to 32 5 inHg 0 1 hPa 0 1 inHg 3 hPa 0 to 60 C 5 hPa 20 to 0 C Battery type CR2 Battery life 72 Days Dimensions 3 x 2 x 1 78 x 50 x 23 mm Weight 2 3 oz 66 g with battery Data points 16000 per parameter ...

Page 15: ...tteries battery leakage or damage resulting from accident tampering misuse or abuse of the product Opening the meter to expose its electronics will void the warranty To obtain warranty service ship the unit postage prepaid to SPER SCIENTIFIC LTD 8281 E Evans Rd Suite 103 Scottsdale AZ 85260 480 948 4448 The defective unit must be accompanied by a description of the problem and your return address ...

Page 16: ...16 rev 8 7 2019 ...