205B Series User’s Manual

 

 118123-001 Rev A 

R

Instruction Manual

Rack-Mount High Voltage Power Supply 

SPELLMAN 

HIGH VOLTAGE ELECTRONICS 

CORPORATION 

One Commerce Park 

Valhalla, New York, 10595 

+1(914) 686-3600 * FAX: +1(914) 686-2870 

E-mail:

[email protected]

Website:

http://www.spellmanhv.com

205B SERIES

Summary of Contents for 205B-01R

Page 1: ...nstruction Manual Rack Mount High Voltage Power Supply SPELLMAN HIGH VOLTAGE ELECTRONICS CORPORATION One Commerce Park Valhalla New York 10595 1 914 686 3600 FAX 1 914 686 2870 E mail sales spellmanhv com Website http www spellmanhv com 205B SERIES ...

Page 2: ...ction table Temperature Coefficient 50ppm C Stability 0 01 hour 0 02 per 8 hours after a 1 2 hour warm up Accuracy Current Monitor 0 5 of reading 0 25 of maximum Remote Programming 0 1 of setting 0 1 of maximum Voltage Monitor 0 1 of reading 0 1 of maximum Front Panel Meter Voltage 0 1 of setting 0 1 of maximum Current 0 25 of setting 0 25 of maximum Front Panel Control 0 25 of setting 0 05 of max...

Page 3: ...488 sold separately 9 Polarity Indicator Open collector 30V 25mA positive ON PAGE 2 OF 2 MODEL SELECTION TABLE 205B Series Voltage Current Ripple 205B 01R 0 to 1kV 0 to 30mA 10mV 205B 03R 0 to 3kV 0 to 10mA 30mV 205B 05R 0 to 5kV 0 to 5mA 50mV 205B 10R 0 to 10kV 0 to 2 5mA 100mV 205B 20R 0 to 20kV 0 to 1mA 300mV 205B 30R 0 to 30kV 0 to 0 5mA 400mV 205B 50R 0 to 50kV 0 to 0 3mA 2 volts DIMENSIONS i...

Page 4: ... aware of the electrical hazards WARNING note in the text call attention to hazards in operation of these units that could lead to possible injury or death CAUTION notes in the text indicate procedures to be followed to avoid possible damage to equipment Copyright 2000 Spellman High Voltage Electronics Corporation All Rights Reserved This information contained in this publication is derived in par...

Page 5: ...spannungsnetzteiles benötigen ca 5 Minuten um sich zu entladen Erden Sie sich nicht und arbeiten Sie nicht in feuchter oder nasser Umgebung Notwendige Reparaturen können es erforderlich machen den Gehäusedeckel während des Betriebes zu entfernen Reparaturen dürfen nur von qualifiziertem eingewiesenem Personal ausgeführt werden WARNING im folgenden Text weist auf gefährliche Operationen hin die zu ...

Page 6: ...de l alimentation Ne vous mettez pas à la masse ou ne travaillez pas sous conditions mouillées ou humides La maintenance peut nécessiter l enlèvement du couvercle lorsque l alimentation est encore allumée Les réparations doivent être effectuées par une personne qualifiée et connaissant les risques électriques Dans le manuel les notes marquées WARNING attire l attention sur les risques lors de la m...

Page 7: ...carica della capacità interna dell alimentatore ad alta tensione Non mettere a terra il proprio corpo oppure operare in ambienti bagnati o saturi d umidità SICUREZZA NELLA MANUTENZIONE Manutenzione potrebbe essere richiesta rimuovendo la copertura con apparecchio acceso La manutenzione deve essere svolta da personale qualificato coscio dei rischi elettrici Attenzione alle AVVERTENZE contenute nel ...

Page 8: ...er 2 2 3 Voltage Control 2 2 4 Polarity Reversal 2 2 4 1 Polarity Reversal 1kV to 5kV Output Models 3 2 4 2 Polarity Reversal 10kV to 50kV Output Models 3 III LOCAL OPERATION 3 1 Front Panel Controls 3 3 2 Program Control Switch 3 3 3 Local Operation 3 IV REMOTE ANALOG OPERATION 4 1 Program Control Switch 3 4 2 Remote Control 3 4 3 Analog Monitoring Signals 3 4 V MAINTENANCE 5 1 General 4 ...

Page 9: ...ly that converts the AC line power to a low voltage DC and a DC to DC converter that generates the high DC voltage Low voltage electronics solid state analog circuits are placed on plug in printed circuit boards The high voltage assembly is fully encapsulated in silicone rubber for reliable arc free stable operation SECTION II OPERATION CAUTION THIS UNIT CAN STORE HAZARDOUS VOLTAGE COMPLETELY DISC...

Page 10: ...the units rear panel PROGRAM CONTROL switch S101 is in the LOCAL position However if the unit receives a voltage programming command while S101 is in the Remote Analog position the switching of S101 to its LOCAL position afterwards will then cause the value of that programming command to take effect 3 3 LOCAL OPERATION The local mode of the 205B Series provides all of the control and monitoring fu...

Page 11: ... collector with respect to digital ground indicating output high voltage polarity NPN saturation denotes positive polarity logic 0 NPN cut off denotes negative polarity logic 1 SECTION V MAINTENANCE 5 1 GENERAL Your 205B Series high voltage power supply is designed for reliable trouble free operation If any questions should arise contact Spellman s Customer Service Department for assistance or ret...

Page 12: ...lleged breach of this warranty No modification amendment supplement addition or other variation of this warranty will be binding unless it is set forth in a written instrument signed by an authorized officer of Spellman Factory Service Procedures For an authorization to ship contact Spellman s Customer Service Department Please state the model and serial numbers which are on the plate on the rear ...