background image

kÉï=~ë=çÑW=

fåëí~ää~íáçå=j~åì~ä

MPKOMNS

loqelmelp=pi=Oa
loqelmelp=pi=Oa=L=`ÉéÜ
loqelmelp=pi=Pa
loqelmelp=pi=Pa=L=`ÉéÜ

båÖäáëÜ

=

båÖäáëÜ

Cover page

Summary of Contents for ORTHOPHOS SL 2D

Page 1: ...k W f j MPKOMNS loqelmelp pi Oa loqelmelp pi Oa L loqelmelp pi Pa loqelmelp pi Pa L b b Cover page...

Page 2: ......

Page 3: ...ns of service personnel 13 2 6 Switching the unit on 13 2 7 Radiation protection 13 2 8 Laser light localizer 13 2 9 Modifications to the unit 14 2 10 Transport safety devices 14 2 11 Electromagnetic...

Page 4: ...1 Standard version 42 5 1 2 Option 1 with second wall holder 44 5 1 3 Option 2 Floor stand installation 45 5 2 Tools materials and measurement tools you will need 47 5 2 1 Tools and materials 47 5 2 2...

Page 5: ...Installation Ceph arm 89 10 1 Installation material 89 10 2 Required tools 90 10 3 Installation 91 10 3 1 Ceph arm mounted on left hand side 91 10 3 2 Ceph arm mounted on right hand side 94 10 3 3 In...

Page 6: ...lect parameters 135 11 12 2 4Saving parameters 135 11 12 2 5Exiting the test step and service routine 136 11 12 3 S017 Configuration service 137 11 12 3 1S017 Test step 2 138 11 12 3 2S017 Test step 3...

Page 7: ...calibration 193 13 1 1 Displays and help messages during adjustment and calibration 194 13 1 2 Calibration menu 195 13 1 2 1 Opening the calibration menu 195 13 1 2 2 Menu structure 197 13 1 3 Enabli...

Page 8: ...Diaphragm adjustment 264 13 2 2 3 Sensor calibration 265 13 2 2 4 Diaphragm configuration 266 13 2 2 5 Geometry calibration 268 13 2 2 6 Dosimetry 270 13 2 3 Resetting adjustment calibration 271 13 3...

Page 9: ...ety information provided in the present operating instructions Such information is highlighted as follows Tip Information on making work easier DANGER An imminent danger that could result in serious b...

Page 10: ...ed The formats and symbols used in this document have the following meaning Prerequisite 1 First action step 2 Second action step or Alternative action Result Individual action step Requests you to do...

Page 11: ...er relevant documents In addition to these installation instructions you will require the following documentation Orthophos SL wiring diagrams Wiring diagrams ORTHOPHOS SL Wiring References REF 64 95...

Page 12: ...sleeves Identification of single use devices Prior to each exposure the hygienic protective sleeves single use devices must be fitted Single use devices are identified with the symbol shown on the lef...

Page 13: ...the X ray tube assembly as the coiled cable of the manual release permits With the exception of the service engineer no other persons are allowed to stay in the room during an exposure In case of mal...

Page 14: ...edical electrical equipment is subject to special EMC related precautions It must be installed and operated as specified in the document Installation Requirements If high voltage systems radio link sy...

Page 15: ...64 95 142 D3632 D3632 031 01 02 02 03 2016 15 Sirona Dental Systems GmbH 3 Unit description Installation Manual ORTHOPHOS SL 3 1 System version b 3 Unit description 3 1 System version A B C...

Page 16: ...THOPHOS SL 3D Digital X ray unit with digital Flat Panel Detector and DCS sensor Flat Panel Detector Digital Flat Panel Detector with a Si technology amorphous silicon Pan sensor Digital CdTe sensor w...

Page 17: ...ion Manual ORTHOPHOS SL 3 2 Sensor versions b 3 2 Sensor versions 2D sensor unit Sensor unit for panoramic X ray PAN DCS sensor 2D 2D 3D sensor unit Sensor unit for panoramic X ray PAN and volume expo...

Page 18: ...31 01 02 02 03 2016 3 Unit description Sirona Dental Systems GmbH 3 3 Dimensions Space requirements Installation Manual ORTHOPHOS SL 3 3 Dimensions Space requirements 3 3 1 Top view A Recommended dist...

Page 19: ...6 19 Sirona Dental Systems GmbH 3 Unit description Installation Manual ORTHOPHOS SL 3 3 Dimensions Space requirements b 3 3 2 Front view A Alternative fastening if it is not possible to screw the unit...

Page 20: ...D3632 031 01 02 02 03 2016 3 Unit description Sirona Dental Systems GmbH 3 3 Dimensions Space requirements Installation Manual ORTHOPHOS SL 3 3 3 Top view with floor stand A Recommended distances fro...

Page 21: ...ona Dental Systems GmbH 3 Unit description Installation Manual ORTHOPHOS SL 3 3 Dimensions Space requirements b 3 3 4 Front view with floor stand A Floor stand Order floor stand separately The unit mu...

Page 22: ...2 D3632 031 01 02 02 03 2016 3 Unit description Sirona Dental Systems GmbH 3 3 Dimensions Space requirements Installation Manual ORTHOPHOS SL 3 3 5 Top view with Ceph left A Recommended distances from...

Page 23: ...rona Dental Systems GmbH 3 Unit description Installation Manual ORTHOPHOS SL 3 3 Dimensions Space requirements b 3 3 6 Front view with Ceph left A Alternative fastening if it is not possible to screw...

Page 24: ...4 D3632 031 01 02 02 03 2016 3 Unit description Sirona Dental Systems GmbH 3 3 Dimensions Space requirements Installation Manual ORTHOPHOS SL 3 3 7 Top view with Ceph right A Recommended distances fro...

Page 25: ...ona Dental Systems GmbH 3 Unit description Installation Manual ORTHOPHOS SL 3 3 Dimensions Space requirements b 3 3 8 Front view with Ceph right A Alternative fastening if it is not possible to screw...

Page 26: ...section Wall mounting standard and option 1 49 B Option 1 Wall mounted installation with 2 wall holders short and no floor fastening if only wall installation is possible on site see section Wall moun...

Page 27: ...mote control with release button on the coiled cable in the treatment room B Installation version 1 Unit with remote control outside the X ray room without release button on the coiled cable see secti...

Page 28: ...are attached to the side of the X ray unit packaging to indicate whether the unit was exposed to a shock during transport White indicator No shock Red indicator Shock Transport and storage temperatur...

Page 29: ...t doesn t necessarily mean that the unit is damaged Make a note on the delivery slip that the indicator is activated Ensure that the driver from the transport company confirms this on the delivery sli...

Page 30: ...designed both for protection during transport and as an installation aid Therefore please remove only the surrounding packaging prior to installation Please leave the Styrofoam packaging and transpor...

Page 31: ...e save the two lateral Styrofoam packaging parts Scope of supply B Filling piece 1 C Filling cardboard 1 D Pack list 1 E Filling cardboard 2 F Filling piece 2 including pack list 2 G Profile cover H I...

Page 32: ...D3632 32 D3632 031 01 02 02 03 2016 4 Delivery and transport Sirona Dental Systems GmbH 4 2 Delivery Installation Manual ORTHOPHOS SL 4 2 2 Sensors DCS sensor A Styrofoam packaging Scope of supply B D...

Page 33: ...D3632 031 01 02 02 03 2016 33 Sirona Dental Systems GmbH 4 Delivery and transport Installation Manual ORTHOPHOS SL 4 2 Delivery b Flat Panel Detector A Styrofoam packaging Scope of supply B Flat Pane...

Page 34: ...0 3 4 x 28 3 4 Weight kg 40 lbs 88 IMPORTANT The right handed ceph arm is packed laterally reversed A Styrofoam packaging Scope of supply B Ceph arm C Accessories 35 and hygienic protective sleeves 35...

Page 35: ...ion Manual ORTHOPHOS SL 4 2 Delivery b 4 2 3 1 Accessories 4 2 3 2 Hygienic protection 20 30 40 50 C A B A Nose support 1x B Ear plug holders with ear plug fixation 2x C Carpus support plate 1x 20 30...

Page 36: ...Systems GmbH 4 2 Delivery Installation Manual ORTHOPHOS SL 4 2 4 Adjustment sets 2D adjustment Panoramic 2D adjustment Cephalometer A A Needle phantom B B Set of Torx offset screwdrivers A C B A Ceph...

Page 37: ...Dental Systems GmbH 4 Delivery and transport Installation Manual ORTHOPHOS SL 4 2 Delivery b 3D adjustment calibration Acceptance and constancy test Worldwide 3D constancy test worldwide A A Geometry...

Page 38: ...02 02 03 2016 4 Delivery and transport Sirona Dental Systems GmbH 4 2 Delivery Installation Manual ORTHOPHOS SL For Germany only 3D constancy test Germany B A A DVT test phantom DIN B Contrast element...

Page 39: ...ackaging attached normal case 1 Open the surrounding packaging using the straps provided A 2 Pull out the carrying handles B 3 Transport the unit to the installation site NOTICE Possible transport dam...

Page 40: ...ral styrofoam packaging parts 4 Loosen the four screws B 5 Pull the carrying handles C out of their holders and insert them through the drillings D of the wooden support A from the inside 6 Insert scr...

Page 41: ...d transport it to the installation site NOTICE Possible transport damage The pan sensor and flat panel detector are sensitive instruments Only remove the packaging immediately prior to installing the...

Page 42: ...016 5 Installation X ray unit Sirona Dental Systems GmbH 5 1 Installation material Installation Manual ORTHOPHOS SL 5 Installation X ray unit 5 1 Installation material 5 1 1 Standard version A For ins...

Page 43: ...all floor mounting Hexagon wood screw 8x80 5 16x3 4 pc Plastic wall plug S10 2 pc Screw M8x30 2 pc Washer 8 4 6 pc M8 nut 2 pc Screw M4x10 18 pc Washer 4 3 2 pc Mounting kit 10 SXR 2 pc TORX offset sc...

Page 44: ...al ORTHOPHOS SL 5 1 2 Option 1 with second wall holder Additional wall holder for bottom wall mounting A For installation on wooden stud frame structures Wall holder long 1 pc Hexagon wood screw 8x80...

Page 45: ...2 02 03 2016 45 Sirona Dental Systems GmbH 5 Installation X ray unit Installation Manual ORTHOPHOS SL 5 1 Installation material b 5 1 3 Option 2 Floor stand installation A Mounting hardware B Floor st...

Page 46: ...al ORTHOPHOS SL Installation with floor stand Floor stand 1 pc Floor stand cover 1 pc Wood screw 10x160 3 8x6 5 pc Plastic wall plug S12 5 pc Screw M8x80 2 pc Washer 8 4 2 pc M8 nut 2 pc Screw M10x50...

Page 47: ...10mm 3 8 Drill hammer drill general purpose Drill or drill hammer depending on the ground Graphics Open end wrench neutral Open end wrench 13 mm A F Open end wrench 13 mm A F Socket wrench Wrench inse...

Page 48: ...wrench Hexagon socket 2 mm Allen key 2 mm Socket wrench Socket wrench and extension 17 Socket wrench and extension Wrench insert 17 5 2 2 Measurement tools Multimeter or ammeter Test unit to measure u...

Page 49: ...cover A 2 Position the two installation aids B at the foot C of the unit and secure their position with adhesive tape CAUTION The installation aids must be placed on top of each other in such a way t...

Page 50: ...thout a pallet set the unit upright with the wooden support You can also use the lateral styrofoam packaging as a support with this variation 4 Loosen the nuts with washers D on both sides of the pall...

Page 51: ...llation X ray unit Installation Manual ORTHOPHOS SL 5 3 Wall mounting standard and option 1 b Mounting the first upper wall holder standard Mount the upper wall holder Mounting the second lower wall h...

Page 52: ...holder must be flush with the upper edge of the unit The unit is now mounted on the upper wall holder 4 Insert screw C through washer D and then through the wall holder and into the stand from behind...

Page 53: ...L 5 3 Wall mounting standard and option 1 b Leveling the unit and fastening it to the floor standard and option 1 1 Remove the transport safety device G 2 Pull wooden board H out of the styrofoam pack...

Page 54: ...lp of the spirit level by adjusting the unit base Align the stand by placing the spirit level on the side and rear of the stand I Check the ring in both directions using the spirit level J 5 Drill thr...

Page 55: ...using the floor stand it must be secured with the upper wall holder NOTICE Reduced image quality Note the wall and ground conditions If a carpet is present at the unit installation site it must be re...

Page 56: ...packaging the two lateral styrofoam parts and the profile cover A 3 Loosen the nuts with washers B on both sides of the pallet or wooden support 4 Take off the pallet or wooden support 5 Remove the th...

Page 57: ...k G NOTICE Do not damage the interfaces or cables Make sure that the interfaces and cables do not bear on the supporting block G and are not damaged when you push the unit 10 Remove the two screws H f...

Page 58: ...L firmly onto the base plate N using the 4 screws and washers M 13 NOTICE Do not damage the cables Ensure that the cables are correctly routed through the support and are not crushed Position the bas...

Page 59: ...ty device 63 2 Insert screw P through the washer and then from the rear through the support and into the stand 3 Fit the profile clamp Q onto the screw P from the other front side and screw the nut R...

Page 60: ...unterclockwise to the front side of the unit 7 Loosen screws P and adjusting nuts O again slightly 8 Level the unit in both directions with the help of the spirit level by adjusting the adjusting nuts...

Page 61: ...3 Attach the unit loosely to the upper wall holder Place the screws U in the grooves Screw the unit onto the wall holder loosely using nuts and washers V NOTICE Do not tighten the screws firmly 4 Leve...

Page 62: ...r stand option 2 Installation Manual ORTHOPHOS SL 5 Drill the fastening holes in the floor through the holes in the base plate Remove the drilling dust with a vacuum cleaner 6 Slide the wall plugs W t...

Page 63: ...in a safe place You will need them should the unit be moved again 5 6 Installing tube assembly cover 1 Loosen screw A on the unit s swivel approx 3 4 turns 2 Fit the Top tube assembly cover B under th...

Page 64: ...holder Installation Manual ORTHOPHOS SL 5 7 Installing the release button holder 1 Unscrew and remove the cover A 2 Attach the holder B 3 Screw the cover A onto the unit B A IMPORTANT Only attach the...

Page 65: ...tools Screwdriver Torx TX10 Screwdriver Torx T10 Screwdriver Torx TX20 Screwdriver Torx TX20 NOTICE Damage to the Flat Panel Detector The Flat Panel Detector is a sensitive component Be careful not t...

Page 66: ...er on the ring Installation Manual ORTHOPHOS SL 6 2 Installing the sensor cover on the ring 1 Loosen screw G approx 2 3 turns 2 Insert the sensor cover H from below over the sensor bracket and then sl...

Page 67: ...lation Manual ORTHOPHOS SL 6 3 Attaching DCS sensor b 6 3 Attaching DCS sensor 1 Carefully remove the DCS sensor from the packaging 2 Fit the DCS sensor into the holes A of the rotation unit and secur...

Page 68: ...g the Flat Panel Detector 1 Carefully remove the Flat Panel Detector from the packaging 2 Fit the Flat Panel Detector from above into the guide of the rotation unit and secure the detector immediately...

Page 69: ...5 Connecting a sensor unit b 6 5 Connecting a sensor unit 1 Insert leads 1 L83A and 2 L83B from the rotation unit as shown in the illustration into plugs X100 X700 and X800 on board DX83 2 If a Flat P...

Page 70: ...5 142 D3632 70 D3632 031 01 02 02 03 2016 6 Installing the sensor unit Sirona Dental Systems GmbH 6 5 Connecting a sensor unit Installation Manual ORTHOPHOS SL 3 Lay the leads as shown in the illustra...

Page 71: ...GmbH 6 Installing the sensor unit Installation Manual ORTHOPHOS SL 6 6 Final installation work b 6 6 Final installation work Connect the sensor covers D and E together over the sensor and secure them...

Page 72: ...ll or drill hammer depending on the ground Graphics Slot screwdrivers assorted Slot screwdriver neutral size Slot screwdriver Torx offset screwdriver graphics Torx offset screwdriver TX20 Torx offset...

Page 73: ...chassis from underneath Preparations 1 Press carefully into the groove A from below with the slotted screwdriver do not lever 2 Remove the lid from the chassis 3 Hold the chassis against the wall in...

Page 74: ...horten cable L117 to the desired length 2 Expose the cable shield D see condition on delivery The length of the wires should be approx 60 mm 2 3 8 3 Place the shortened shielding over the insulation a...

Page 75: ...mote control Installation Manual ORTHOPHOS SL 7 3 Installation b Attaching the strain relief Left concealed installation right surface installation Attach cable L117 to the strain relief in the chassi...

Page 76: ...to the board 4 Connect control cable L117 to terminal X108 board DX42 as shown in the connection diagram Shorten control cable L117 as required 5 Lay cable L117 in a loop following the bottom edge of...

Page 77: ...the rear 2 Using the two screws C fasten the release button holder B to the keyboard 3 Plug the connector of the coiled cable into socket X101 on board DX42 and screw the connector down tight 4 Hook...

Page 78: ...surface installation 2 Attach the cable to the still exposed strain relief 3 Connect the current carrying cables and the ground wire to terminal X105 board DX42 as shown in the connection diagram 4 Se...

Page 79: ...ORTHOPHOS SL 7 6 Final installation work b 7 6 Final installation work 1 Reassemble the remote control 2 NOTICE Make sure that no cables are pulled off when you clip the cover on Hang the release but...

Page 80: ...S SL 7 7 Connecting a door contact switch Connect the door contact switch between terminal X108 pin 1 board DX42 and control cable L117 pin 1 BU X108 1 2 3 4 5 6 Pin WHGN GN BN WHBN BU WHOG WHBU OG X1...

Page 81: ...on version without remote control Connect the coiled cable of the release button to socket X103 of the unit Installation version with remote control 1 Use cable L117 to connect the remote control to s...

Page 82: ...ing demo mode operation without radiation release DANGER Perilous shock hazard Switch off the power supply connection before connecting the line voltage DANGER Perilous shock hazard Fixed connection I...

Page 83: ...leakage current according to IEC 62353 2007 see sections Checking the protective ground wires 84 and Checking the leakage current of the unit 88 2 Record the measured values in the document titled Ins...

Page 84: ...on 2 Disconnect the power cable and the second protective ground wire from the building installation 3 Remove the bottom profile front tube assembly and rear tube assembly covers Measuring setup for p...

Page 85: ...ng the formula R U I If the resistance value is greater than indicated in the adjacent table check whether the protective ground wires are fastened according to the specifications Check whether plain...

Page 86: ...64 95 142 D3632 86 D3632 031 01 02 02 03 2016 9 Safety checks Sirona Dental Systems GmbH 9 1 Checking the protective ground wires Installation Manual ORTHOPHOS SL Measuring point A central ground wire...

Page 87: ...ORTHOPHOS SL 9 1 Checking the protective ground wires b Measuring points B and C GNYE power connection and 2nd protective ground wire Measuring points D and E Board cage DX32 and tube assembly housing...

Page 88: ...instructions of the tester 7 Document the measured value of the leakage current in the technical document Inspection and maintenance and safety related checks REF 64 95 100 to identify changes from th...

Page 89: ...h arm 10 1 Installation material NOTICE Reduced image quality For operation of the unit with floor stand and Ceph an additional mounting using the upper wall holder is absolutely necessary No Ceph ins...

Page 90: ...ols Graphics Open end wrench neutral Open end wrench 8 mm A F Open end wrench 8 mm A F Open end wrench 10 mm A F Open end wrench 10 mm A F Graphics Neutral Allen wrench Allen wrench 2 5 Allen wrench 2...

Page 91: ...A 2 CAUTION Do not yank out the ambient light cable Remove the cover B Tip You can even leave the cover B on the unit and push it upwards a little 3 Remove the cover C by pulling it upwards 4 Reattach...

Page 92: ...ofoam packaging on top 3 Remove the adhesive tape and the cover E from the ceph packaging 4 Break off the wall part F of the packaging at the perforation 5 Slide the cardboard support with the ceph ar...

Page 93: ...ame height as the drill holes on the ceph support arm NOTICE Check the room height before you raise the unit If the room height is less than 2 27 m 89 3 8 or 2 30 m 90 1 2 for installation with the fl...

Page 94: ...t 1 Loosen the screws A 2 CAUTION Do not yank out the ambient light cable Remove the cover B Tip You can even leave the cover B on the unit and push it upwards a little 3 Remove the cover C by pulling...

Page 95: ...o the wall so that the drill holes of the support arm are horizontally aligned with the positions of the threaded bolts 6 Switch on the unit G During the unit startup phase the start screen appears on...

Page 96: ...keys I on the control panel to lower the unit until the bearing bolts on the unit are at the same height as the drill holes on the ceph support arm NOTICE Check the room height before you raise the un...

Page 97: ...n them firmly 3 Detach the cables from the holder Lay the cables over the support arm in the groove between the slide and the support arm from left to right toward the opening K 4 NOTICE Never crush t...

Page 98: ...ragm 1 Remove the secondary diaphragm from the packaging 2 Remove the safety cable tie A for the screws 3 Guide the secondary diaphragm beneath the ceph arm and slide the connector upward through the...

Page 99: ...2 Screw in place the plug connections L35 L38 3 Roll the plug connection L36 L39 into a loop and secure it with a cable tie 4 Grasping inside the ceph arm tube from below carefully pull cable L35 two...

Page 100: ...02 02 03 2016 10 Installation Ceph arm Sirona Dental Systems GmbH 10 4 Connecting control cables Installation Manual ORTHOPHOS SL 5 Roll the cable into a loop and stow it away in the tube 6 Close the...

Page 101: ...uide the cable on the unit through the groove A 2 Lay the cable and the cable loop L36 L39 as illustrated 3 NOTICE Ensure that you do not crush or kink the cables Attach the cover B 10 4 2 Running the...

Page 102: ...the secondary diaphragm 2 NOTICE Do not damage the unit Do not move the secondary diaphragm or sensor adapter by hand or apply any force Remove the top cover 3 Remove the protective canvas A 4 Loosen...

Page 103: ...ystems GmbH 10 Installation Ceph arm Installation Manual ORTHOPHOS SL 10 5 Final installation work b 8 Insert both ear plug holders and fasten them securely using the screws E 9 Press the button F and...

Page 104: ...ers 1 For installation with floor stand only Place the four spring clips E on the base plate The two covers F and G must hold together after they have been attached 2 Attach the profile cover J To do...

Page 105: ...ensor for ceph versions NOTICE Ensure that the correct sensor for cephalometry ceph sensor is plugged in Insert the ceph sensor A carefully upward into the holder until it audibly clicks into place 11...

Page 106: ...e when it moves exception Calibration block and positioning aids NOTICE Fluctuations in temperature can cause condensation to form in the unit Electrical components are destroyed by short circuits Do...

Page 107: ...lights up This LED remains lit as long as the unit is on The start screen is displayed on the touchscreen for several seconds The program selection is then displayed on the touchscreen The forehead su...

Page 108: ...ration on delivery The acoustic signal for end of exposure is activated English is preconfigured as the unit language The preview image is switched on The welcome screen is switched on The first name...

Page 109: ...e test cycle or the rotation unit stops due to increased resistance the assembly of the sensor must be checked again 11 6 1 Starting the test cycle The test cycle is executed without radiation The tes...

Page 110: ...move the unit up and down through its entire adjustment range 2 Check the height adjustment mechanics Does the height adjustment function correctly Are there any unusual noises during the adjustment...

Page 111: ...service routine S018 2 Unit variants For unit orders with the ceph arm the unit version must be configured through service routine S017 2 138 Remote control All units ordered with or without remote co...

Page 112: ...ssigned a unique IP address and the name of the X ray component must also be unique Factory setting of the TCP IP address of the unit 192 168 15 240 subnet mask 255 255 255 0 NOTE The network configur...

Page 113: ...an X ray component SIDEXIS 4 is installed The Sirona AcquisitionServer software is installed The workstation plug in software for the corresponding X ray component or sensor is installed on the workst...

Page 114: ...ray components 4 If necessary Select the acquisition server from the Acquisition Server list field 5 Click on the Next button A password dialog box appears 6 Enter the service password and confirm you...

Page 115: ...available X ray components The X ray components found on the network are displayed in list form NOTE If no X ray component is found check the availability of the unit and repeat the Automatic search...

Page 116: ...ice Administration The X ray component appears in the device list of the configuration menu ORTHOPHOS SL Device Administration 6 Click on the Cancel button 7 Reboot the SIDEXIS 4 8 Register a patient...

Page 117: ...ens Boot service 1 Click the Boot service button in the structure tree 2 A message window prompts you to switch the new X ray component back on to reboot The network is searched for the booting X ray...

Page 118: ...st the network configuration IP address subnet mask and standard gateway to your network 4 Click the Save button to save the settings 5 Wait until the saving operation has been completed NOTICE If you...

Page 119: ...vice Administration The X ray component appears in the device list of the configuration menu ORTHOPHOS SL Device Administration 7 Click on the Cancel button 8 Reboot the SIDEXIS 4 9 Register a patient...

Page 120: ...anually The menu for adding X ray components opens Add device manually 1 Click the Add device manually button in the structure tree 2 Enter the IP address and the location of the desired X ray compone...

Page 121: ...vice Administration The X ray component appears in the device list of the configuration menu ORTHOPHOS SL Device Administration 6 Click on the Cancel button 7 Reboot the SIDEXIS 4 8 Register a patient...

Page 122: ...mo mode and repacking Sirona Dental Systems GmbH 11 9 Checking the data paths Installation Manual ORTHOPHOS SL 11 9 Checking the data paths Checking the 2D and 3D data paths is possible through the se...

Page 123: ...ernet Explorer or Firefox 2 In the address line enter https and the IP address of your unit Example https 192 168 15 240 NOTICE The IP address of your unit can be found on the Info screen Calling the...

Page 124: ...na Dental Systems GmbH 11 10 Setting ambient light via the Web service Installation Manual ORTHOPHOS SL 5 Change the color by clicking the color scale and brightness by clicking on the brightness scal...

Page 125: ...Set dip switch S2 DX6 to position 2 IMPORTANT If switch S2 is not set to position 2 in demo mode before switching off the unit various error messages will display when the unit is turned back on 5 Pu...

Page 126: ...connector J6 J103 DX6 Radiation release is now once again possible 6 Switch on the unit and check the mode on the info screen Demo mode ON means that Demo mode is switched on radiation release is not...

Page 127: ...repacking and transport Packing the ORTHOPHOS SL IMPORTANT If a used carton on which one of the shockwatch or tiltwatch indicators has already been tripped is used to package the unit please make an...

Page 128: ...assembly guard supplied in the package to the X ray tube assembly and secure the packing position of the tube assembly with the tension belt supplied in the package 5 Attach the transport safety devi...

Page 129: ...utines S008 3 S017 2 S017 18 and S037 4 can also be run via the Sirona web service see section Running service routines via the web service 174 Service routines S017 and S037 must sometimes be perform...

Page 130: ...OS SL 11 12 1 Service menu and service routines You can use the service routines to check the function of certain unit components and modules as well as to set important unit parameters 11 12 1 1 Disp...

Page 131: ...and arrow keys to select unit parameters in the selection fields F S1 S37 Selected service routine G 1 n Selected test step H Patient symbol keys Different functions depending on service routine I R...

Page 132: ...ines can be activated and unit settings incl tests and calibrations can be performed in this menu The touchscreen display is located on level 1 1 Touch the toothed wheel A in the upper right corner of...

Page 133: ...a the arrow keys in the selection field 1 A and confirm the selection via the service key B If the selected service routine has several test steps the first selectable test step is displayed in select...

Page 134: ...prevents the inadvertent selection or activation of these service routines To select a service routine or test step with security access proceed as follows 1 Select the service routine or the test ste...

Page 135: ...ce routine S017 6 the code 01 is pre selected for the Remote control enabled option Press the or arrow keys A to select code 00 B for the Remote control disabled option Once the selected parameter has...

Page 136: ...12 Service routines for commissioning and repacking Installation Manual ORTHOPHOS SL 11 12 2 5 Exiting the test step and service routine Touch the Service key A or the double arrow key B to go back to...

Page 137: ...l s touchscreen Additional service routines can be performed Before a patient exposure can be performed the error message must be acknowledged with the R button and the unit must be restarted SR SA Fu...

Page 138: ...on field 1 2 Use arrow keys A to select the required unit variants in selection field 1 see table Once the unit variant has been selected the Memory key B lights up 3 Save the setting 4 Exit the servi...

Page 139: ...e routines for commissioning and repacking b Checking the jumper position on board DX1 Check the jumper positions of the sockets on board DX1 X309 DX91 ceph sensor and X503 DX81 ceph sensor Without ce...

Page 140: ...country group code Factory setting 1 Call service routine S017 3 Once the service routine has been selected the code for the current setting is displayed in selection field 1 2 Use arrow keys A to se...

Page 141: ...4 Once the service routine has been selected the code for the current setting is displayed in selection field 1 2 Use the arrow keys A to select the code for the required language in selection field 1...

Page 142: ...e control display factory settings 1 Call service routine S017 6 Once the service routine has been selected the code for the current setting is displayed in selection field 1 2 Use the arrow keys A to...

Page 143: ...y setting 1 Call service routine S017 8 Once the service routine has been selected the code for the current setting is displayed in selection field 1 2 Use arrow keys A to select the required setting...

Page 144: ...come screen Factory setting 1 Call service routine S017 13 Once the service routine has been selected the code for the current setting is displayed in selection field 1 2 Use arrow keys A to select th...

Page 145: ...de for the line currently selected is displayed in selection field 1 2 Use arrow keys A to select the required line in selection field 1 see table The activation status code is displayed in selection...

Page 146: ...re 1 Call service routine S017 15 Once the service routine has been selected the code for the current setting is displayed in selection field 1 2 Use arrow keys A to select the required setting in sel...

Page 147: ...l angle deviating from the factory setting 7 make the corresponding configuration settings via service routine S018 7 factory settings 1 Call service routine S017 18 147 Once the service routine has b...

Page 148: ...tment S018 2 No Set the maximum travel height S018 3 No Undo the maximum travel height setting S018 4 No Check the height adjustment sensor system S018 5 No Setting the minimum travel height S018 6 No...

Page 149: ...user mode on the control panel 2 Call service routine S018 2 Once the service routine has been selected the current height position is displayed in selection field 1 The Memory key B lights up 3 To s...

Page 150: ...RTHOPHOS SL 11 12 4 2 S018 Test step 3 Undo the maximum travel height setting 1 Call service routine S018 3 Once the service routine has been selected the current height position is displayed in selec...

Page 151: ...een selected the current height position is displayed in selection field 1 Patient symbol keys 1 to 3 F show the switching state of the limit switches see table If the RESPECTIVE patient symbol key is...

Page 152: ...n selected the current height position is displayed in selection field 1 The Memory key B lights up 3 To save the minimum travel height touch the Memory key B R key C lights up followed by the R key C...

Page 153: ...e nominal value displayed in selection field 2 applies The nominal value for the program group selected in selection field 1 is displayed in selection field 2 The currently measured angle of the bite...

Page 154: ...ible range 17 to 22 Tip You can also set the required angle manually on the occlusal bite block and transfer this angle to selection field 2 by pressing the T key J By pressing patient symbol key 1 G...

Page 155: ...splayed in selection field 1 The currently measured AD value leading to the angle calculation is displayed in selection field 2 The codes of the possible error modes of the acquisition electronics are...

Page 156: ...onnections on board DX1 and occlusal bite block electronics 04 Interruption of the 5 V line 05 Interruption of Vref line 06 Interruption of Vout line 07 Interruption of VGND line 08 Interruption of Co...

Page 157: ...gged in and the bite block plate is deflected all the way down Call service routine S018 9 The data type intended for the next storage function is displayed in selection field 1 The display of patient...

Page 158: ...st press the Memory key the R key lights up and then the R key The value for the upper value U is accepted L Adjusting and saving the lower value 1 Deflect the bite block plate all the way down up to...

Page 159: ...e block must be deflected in such a way that the creep speed is effective in the relevant direction observe green arrows The PWM index values are set correctly if the drive runs at approximately the s...

Page 160: ...10 The PWM index for upward travel U up is displayed in selection field 1 The PWM index for downward travel D down is displayed in selection field 2 The display of patient symbol key 1 G indicates wh...

Page 161: ...l service routine S021 3 133 An inactive progress indicator in selection field 1 and the message FFFF in selection field 2 signal that the system is ready to move the diaphragm 2 Touch the T key A The...

Page 162: ...cking position for the sensor unit 1 Call service routine S021 4 133 An inactive progress indicator in selection field 1 and the message FFFF in selection field 2 signal that the system is ready to mo...

Page 163: ...step 10 Self test of the DCS image data path 1 Call service routine S032 10 133 Once the service routine has been selected DCS test is displayed in selection field 1 2 Press the T key The DCS image da...

Page 164: ...tallation Manual ORTHOPHOS SL 11 12 6 2 S032 Test step 50 Self test of the FP image data path 1 Call service routine S032 50 133 Once the service routine has been selected FP test is displayed in sele...

Page 165: ...nsmissions and CAN bus commands The image data are analyzed in the memory of the DX11 It is possible to evaluate the LEDs on DX1 image pan during the synchronized readout sequence Image LED V501 Pan L...

Page 166: ...Replace sensor unit If error messages are present please implement the measures according to the displayed error message Transfer Minimal image data was received by the sensor unit during the self tes...

Page 167: ...the default IP address default gateway address and default subnet mask S037 3 yes Configuring boot mode DYNAMIC DHCP AutoIP STATIC fixed address S037 4 yes Manual input of static network settings IP...

Page 168: ...vice routine S037 1 Once the service routine has been selected the IP address of the unit is displayed in selection field 1 see table Default or static or dynamic is displayed in selection field 2 see...

Page 169: ...r to the restore operation see table The patient symbol key selected in each case lights up 3 To reset the network data press the Memory key E R key lights up followed by the R key F The default netwo...

Page 170: ...g 1 Call service routine S037 3 Once the service routine has been selected the current boot mode of the unit is displayed in selection field 1 2 Use arrow keys A to select the required boot mode autom...

Page 171: ...lit The IP address is displayed in selection field 1 or after pressing the T key number pad B1 is selected Patient symbol key 2 B lit The default gateway is displayed in selection field 1 or after pre...

Page 172: ...field 1 by pressing the patient symbol keys A B or C see table The patient symbol key selected in each case lights up 3 To change the selected parameter first press the T key E 4 Now use the patient...

Page 173: ...b 5 Use the arrow keys F to select the digit to be changed in selection field 1 Digit No 12 in the example The corresponding patient symbol key D lights up Selection field 2 displays the value of the...

Page 174: ...e S017 6 Activating deactivating the remote control S017 8 Selecting the kV mA level series S017 13 S017 14 Enable disable the welcome screen or the lines in the welcome screen S017 15 Activate deacti...

Page 175: ...site The Sirona Web service is loaded The sub menus of the User menu are displayed in the structure tree 5 Click on the Service button in the header line of the main menu The dialog box for entering t...

Page 176: ...o the service routines that can be called using the control panel see chapter Perform service routines via the control panel 9 Enter the desired parameters in the service menu and confirm the entries...

Page 177: ...s or remove a cover For measurements in the area of the power supply terminal the unit must be disconnected from the junction box of the building installation before you start to connect test cables D...

Page 178: ...y should be switched off at the junction box of the building installation when performing measurements in the vicinity of the power supply unit Select the correct current voltage type and adjust the m...

Page 179: ...lsed radiation of type Mult O Meter 512L or similar Measuring wires Measuring wires with the following properties dielectric strength 1000V CAUTION Only use a battery operated digital multimeter with...

Page 180: ...t at least 1 minute before starting to check the line voltage Disconnect the unit from the junction box of the building installation 2 Remove the Bottom profile cover see Service Manual 3 CAUTION Only...

Page 181: ...he Installation Requirements must be installed If the measured voltage drop falls within the permissible tolerance range see the following table finalize the measurement Permissible voltage drop Concl...

Page 182: ...ready for exposure in SIDEXIS 4 see Operating Instructions 4 CAUTION Activating the release button triggers X rays Start the exposure by pressing the release button Hold down the release button until...

Page 183: ...he kV mA level 60 kV 8 mA 7 Use the arrow keys A in selection field 2 to select the radiation time 0 5 s Performing measurements Initiate the radiation Hold the release button pressed until the set ra...

Page 184: ...the unit wait at least 4 minutes LED V500 on the DX6 must no longer be on before removing the cover on the tube assembly 2 Loosen the screw A and remove the lid of the tube assembly cover B 3 Loosen t...

Page 185: ...64 95 142 D3632 D3632 031 01 02 02 03 2016 185 Sirona Dental Systems GmbH 12 Startup measurements and tests for USA Canada only Installation Manual ORTHOPHOS SL 12 7 Checking the tube current b...

Page 186: ...DANGER Do not touch any live components Set the main switch A to I see also Operating Instructions 9 Wait for approx 1 minute 10 Press the R key The unit moves to its starting position Performing meas...

Page 187: ...ng Instructions 2 Loosen the screw and remove the lid of the tube assembly cover 3 DANGER After switching off the unit wait at least 4 minutes before removing the measuring wires or reinserting the ju...

Page 188: ...localizers Checking the laser light localizers Preparing the test 1 Insert the forehead and temple supports 2 Set the main switch A to I see also Operating Instructions 3 Wait for approx 1 minute 4 T...

Page 189: ...ight localizer is switched back off again Checking the 3D laser beam 1 Touch the 3D symbol at the top of the touchscreen The 3D program group is selected 2 Select the exposure program Press the arrow...

Page 190: ...ed 6 Press on the light localizer key again The light localizer is switched back off again Checking the horizontal laser beam ceph FH Tip When checking or adjusting the light localizer you may use a P...

Page 191: ...rforming the following tests be sure to observe the radiation protection regulations applicable in your country see Operating Instructions DANGER X rays Radiation is signaled by the message X RAY acti...

Page 192: ...64 95 142 D3632 192 D3632 031 01 02 02 03 2016 13 Unit adjustment and calibration Sirona Dental Systems GmbH Installation Manual ORTHOPHOS SL A...

Page 193: ...amic Sensor adjustment Diaphragm adjustment Symmetrie adjustment Sensor calibration DCS If a cephalometer is installed Adjustment of Ceph primary diaphragm Adjustment of Ceph secondary diaphragm Adjus...

Page 194: ...or exposure H4 04 Plug the sensor into the ceph slot incorrect sensor plugged in Status messages Ready for exposure Exposure not possible Please wait Ready for exposure in XXs X RAY Active Exposure is...

Page 195: ...the unit 13 1 2 1 Opening the calibration menu IT package is installed is installed 1 Start SIDEXIS 4 2 In SIDEXIS 4 call the ORTHOPHOS SL Device administration configuration menu Configuration menu O...

Page 196: ...ystem date PC in reverse order Example On 25 03 2016 the service password is 3052 Service mode When you open the calibration menu the unit switches from user mode to the PC service mode blocked by the...

Page 197: ...16 197 Sirona Dental Systems GmbH 13 Unit adjustment and calibration Installation Manual ORTHOPHOS SL 13 1 General information on unit adjustment and calibration b 13 1 2 2 Menu structure The menu is...

Page 198: ...ta entry and display window C Preview image Shows the exposure to be taken in this stage of the adjustment calibration procedure D Tools pictograph Shows which if any test phantom must be used for thi...

Page 199: ...ifferent areas the menu also contains the following buttons I Fine adjustment Before the fine adjustment and with the check box unchecked a coarse adjustment is performed 203 J Start acquisition Creat...

Page 200: ...GmbH 13 1 General information on unit adjustment and calibration Installation Manual ORTHOPHOS SL Toolbar functions Button Function Rotate exposure 90 clockwise Reducing size of section Increasing si...

Page 201: ...orm adjustment and calibration of the unit in the order as described in the section General information on unit adjustment and calibration 193 Structure tree Resetting adjustment calibration and data...

Page 202: ...are prefixed by status indicators indicating the current status of the corresponding adjustment or calibration operation Adjustment Calibration Not checked Invalid data record or no record present Ad...

Page 203: ...on adjustment is preset Always try to use precision adjustment first when adjusting the unit In most cases previous coarse adjustment is not necessary A message window indicates that a coarse adjustme...

Page 204: ...or adjustment can usually be performed directly via precision adjustment Only in exceptional cases e g if one or several needles are completely outside of the image section in an image acquired with t...

Page 205: ...sually be performed directly via precision adjustment Only in exceptional cases e g if the exposed image area is completely outside the image section in an image acquired with the Fine adjustment sett...

Page 206: ...s the ceph secondary diaphragm can be performed directly via precision adjustment Only in exceptional cases e g if the exposed image area is completely outside the image section in an image acquired w...

Page 207: ...f the exposed image section in the direction of the stationary auxiliary lines see the example Example In the example the exposed image section is offset to the left by the value Sx and upward by the...

Page 208: ...matic adjustment until the default values in the text boxes are equal to zero This means that adjustment has been successful If the values do not converge toward zero despite repeating the adjustment...

Page 209: ...on the control panel along with the specific exposure parameters 13 1 4 Taking an exposure The calibration menu must be opened Select the corresponding element in the structure tree The unit must be...

Page 210: ...justments the needle phantom A must be inserted into the pan bite block holder of the unit For Diaphragm adjustment panorama adjustment the needle phantom must be taken out of the bite block holder NO...

Page 211: ...nt and calibration Installation Manual ORTHOPHOS SL 13 1 General information on unit adjustment and calibration b When fitting the needle phantom make sure that it is correctly oriented For the panora...

Page 212: ...OPHOS SL 13 1 6 2 Adjustment phantom for adjustment of the cephalometer For Adjustment of Ceph primary diaphragm and Adjustment of Ceph main beam direction ceph adjustments the ceph adjustment phantom...

Page 213: ...ometry phantom A into the pan bite block holder B on the unit For Diaphragm adjustment 3D adjustment as well as for Sensor calibration DCS and Diaphragm calibration 3D calibrations you must take the g...

Page 214: ...Installation Manual ORTHOPHOS SL 1 Insert the geometry phantom A into the pan bite block holder B on the unit and align the geometry phantom correctly with the screw C 2 Attach the clip D for better s...

Page 215: ...t and calibration via the calibration menu Only for Germany 13 2 1 2D adjustment calibration 13 2 1 1 Pan sensor adjustment IMPORTANT After every adjustment and calibration of the unit the reference v...

Page 216: ...ure tree click on the Sensor adjustment element S010 1 under Pan The Sensor adjustment menu is displayed in the action area The menu supports precision adjustment and coarse adjustment precision adjus...

Page 217: ...ting at Step 3 until the values in the menu s text boxes are equal to zero This means that adjustment was performed successfully IMPORTANT If automatic adjustment fails to produce values equal to zero...

Page 218: ...ibration menu Installation Manual ORTHOPHOS SL 6 If adjustment is OK the adjustment values in the text boxes are equal to zero save the values 209 The status indicator that prefixes the Sensor adjustm...

Page 219: ...ed in the chapter entitled Automatic adjustment Panoramic sensor standard 216 2 After taking the exposure measure the distances S1 S2 and S3 with the toolbar measuring ruler 197 To measure S1 S2 and S...

Page 220: ...ustment values for S1 S2 and S3 in the text boxes of the Sensor adjustment menu with the measured values For information about the shifting direction input of sign in the menu see the chapter entitled...

Page 221: ...the diaphragm is tilted appears in the menu during Diaphragm adjustment the primary diaphragm is mechanically maladjusted in the vertical axis In this case adjust the diaphragm 237 and repeat after t...

Page 222: ...ement S030 2 The Diaphragm adjustment menu is displayed in the action area The menu supports precision adjustment and coarse adjustment precision adjustment is pre selected Perform a precision adjustm...

Page 223: ...64 95 142 D3632 D3632 031 01 02 02 03 2016 223 Sirona Dental Systems GmbH 13 Unit adjustment and calibration Installation Manual ORTHOPHOS SL 13 2 Adjustment and calibration via the calibration menu b...

Page 224: ...zero calculate the adjustment values manually and overwrite the values in the menu s text boxes with these values 225 5 If adjustment is OK the adjustment values in the text boxes are equal to zero s...

Page 225: ...adjustment The only difference is that the default adjustment values calculated automatically in the Diaphragm adjustment menu are overwritten by adjustment values calculated manually 1 Start the Diap...

Page 226: ...justment values for Sx and Sy in the text boxes of the Diaphragm adjustment menu with the measured values For information about the shifting direction input of sign in the menu see the chapter entitle...

Page 227: ...mmetry The needle phantom is in the bite block holder on the unit 210 1 In the structure tree under Pan click on the Symmetrie adjustment element S010 2 The Pan Symmetrie menu is displayed in the acti...

Page 228: ...64 95 142 D3632 228 D3632 031 01 02 02 03 2016 13 Unit adjustment and calibration Sirona Dental Systems GmbH 13 2 Adjustment and calibration via the calibration menu Installation Manual ORTHOPHOS SL...

Page 229: ...u s text boxes with these values 230 5 If adjustment is OK the adjustment values in the text boxes are equal to zero save the values 209 The status indicator that prefixes the Symmetrie adjustment ele...

Page 230: ...pter entitled Automatic adjustment Pan Symmetry 227 2 After taking the exposure measure the distances S1 S2 and S3 with the toolbar measuring ruler 197 If possible measure in the lower area of the nee...

Page 231: ...n the structure tree under Pan click on the Sensor calibration DCS element S010 10 3 Establish 209 receptivity 4 Take the exposure for the first kV mA combination 60kV 5mA 15 1s 209 After taking the e...

Page 232: ...ombinations After the last exposure has been taken a message in the action window indicates whether calibration has been successful and the status indicator at the start of the element Sensor calibrat...

Page 233: ...m connects to the cover of the sensor unit 212 The ceph sensor is plugged into the ceph slot on the unit The ear plug holders on the ceph arm have been moved completely apart and swung out of the beam...

Page 234: ...d calibration via the calibration menu Installation Manual ORTHOPHOS SL 3 Create an exposure 64 kV 16 mA 6 1 s 209 The adjustment values S1 S2 S3 and S4 are automatically determined from the exposure...

Page 235: ...3632 D3632 031 01 02 02 03 2016 235 Sirona Dental Systems GmbH 13 Unit adjustment and calibration Installation Manual ORTHOPHOS SL 13 2 Adjustment and calibration via the calibration menu b A B S2 S2...

Page 236: ...menu s text boxes with these values 237 5 If adjustment is OK the adjustment values in the text boxes are equal to zero save the values 209 The status indicator that prefixes the Adjustment of Ceph pr...

Page 237: ...in the chapter entitled Automatic adjustment Ceph Primary diaphragm standard 233 2 Once you have taken the exposure measure distances S1 S2 S3 and S4 with the toolbar measuring ruler 197 S1 distance...

Page 238: ...in the pan bite block holder on the unit The ceph test phantom does not connect to the cover of the sensor unit IMPORTANT If you cannot correctly adjust the ceph secondary diaphragm via the calibrati...

Page 239: ...diaphragm menu is displayed in the action area The menu supports precision adjustment and coarse adjustment precision adjustment is pre selected Perform a precision adjustment first In most cases prev...

Page 240: ...tion via the calibration menu Installation Manual ORTHOPHOS SL A Adjustment OK The exposed diaphragm area must lie centered and straight in the image section as well as inside the superimposed auxilia...

Page 241: ...o This means that adjustment was performed successfully IMPORTANT If automatic adjustment fails to produce values equal to zero calculate the adjustment values manually and overwrite the values in the...

Page 242: ...hragm The manual adjustment procedure is similar to the one for automatic adjustment 238 The only difference is that the default adjustment values calculated automatically in the Adjustment of Ceph se...

Page 243: ...ntal Systems GmbH 13 Unit adjustment and calibration Installation Manual ORTHOPHOS SL 13 2 Adjustment and calibration via the calibration menu b 2 Once you have taken the exposure measure distances S1...

Page 244: ...for S1 S2 and S3 in the text boxes of the Adjustment of Ceph secondary diaphragm menu with the measured values For information about the shifting direction input of sign in the menu see the section en...

Page 245: ...d calibration Installation Manual ORTHOPHOS SL 13 2 Adjustment and calibration via the calibration menu b 13 2 1 6 3 Mechanical adjustment Ceph secondary diaphragm 1 Loosen screw A and remove the cove...

Page 246: ...rn counterclockwise adjusts the diaphragm to the right Turn clockwise adjusts the diaphragm to the left Screw D Turn counterclockwise adjusts the diaphragm downward Turn clockwise adjusts the diaphrag...

Page 247: ...an needle phantom is not in the pan bite block holder on the unit The ceph test phantom does not connect to the cover of the sensor unit 1 In the structure tree under Ceph click on the Adjustment of C...

Page 248: ...nt and calibration via the calibration menu Installation Manual ORTHOPHOS SL 3 Create an exposure 80kV 14 mA 0 3 s 209 Adjustment value S1 is calculated automatically from the exposure and entered in...

Page 249: ...djustment values manually and overwrite the values in the menu s text boxes with these values 242 5 If adjustment is OK the adjustment values in the text boxes are equal to zero save the values 209 Th...

Page 250: ...nual adjustment procedure is similar to the one for automatic adjustment 247 The only difference is that the default adjustment values calculated automatically in the Adjustment of Ceph secondary diap...

Page 251: ...Sirona Dental Systems GmbH 13 Unit adjustment and calibration Installation Manual ORTHOPHOS SL 13 2 Adjustment and calibration via the calibration menu b 2 Once you have taken the exposure measure di...

Page 252: ...S1 in the text box of the Adjustment of Ceph secondary diaphragm Quickshot menu with the measured value For information about the shifting direction input of sign in the menu see the chapter entitled...

Page 253: ...connects to the unit s sensor unit 212 The ear plug holders on the ceph arm have been swung out of the beam direction ap 1 In the structure tree under Ceph click on the Adjustment of Ceph main beam di...

Page 254: ...al Systems GmbH 13 2 Adjustment and calibration via the calibration menu Installation Manual ORTHOPHOS SL 3 Create an exposure 80kV 14mA 14 9 s 209 Adjustment value S1 is calculated automatically from...

Page 255: ...es equal to zero calculate the adjustment values manually and overwrite the values in the menu s text boxes with these values 256 5 If adjustment is OK the adjustment values in the text boxes are equa...

Page 256: ...matic adjustment Ceph Main X ray beam direction standard 253 2 Once you have taken the exposure measure the distance S1 with the toolbar measuring ruler 197 Measure within the measuring range shown in...

Page 257: ...earplug alignment 13 2 1 9 1 Checking the ear plug alignment The ceph sensor is plugged into the ceph slot on the unit 1 In the structure tree under Ceph click on the Adjustment of earplug alignment...

Page 258: ...ide sensor side Transparent adjusting cap B on the inside tube assembly side 4 Establish 209 receptivity 5 Create an exposure 80kV 14mA 14 9 s 209 The exposure is displayed in the exposure window The...

Page 259: ...na Dental Systems GmbH 13 Unit adjustment and calibration Installation Manual ORTHOPHOS SL 13 2 Adjustment and calibration via the calibration menu b 13 2 1 9 2 Adjusting the ear plug alignment Unscre...

Page 260: ...o not unscrew them fully 2 Adjust the ear plugs in the horizontal direction by turning the screw B counterclockwise or clockwise The direction of rotation for left arm C and right arm D versions is sh...

Page 261: ...ibration menu b 3 Adjust the ear plugs in the vertical direction by turning the knurled nut F The direction of rotation for left arm G and right arm H versions is shown in the drawing below 4 Tighten...

Page 262: ...2 3D adjustment calibration 13 2 2 1 Sensor adjustment 1 Insert the 3D calibration block into the bite block holder on the unit 213 IMPORTANT For the 3D adjustment and calibration the values in the in...

Page 263: ...element S010 20 5 Establish readiness for exposure 209 6 Create an exposure 85kV 22 mAs 209 A message is displayed in the action window to indicate whether adjustment has been successful and the value...

Page 264: ...S030 10 2 Establish readiness for exposure 209 3 Create an exposure 85kV 0 34 mAs 209 A message is displayed in the action window to indicate whether adjustment has been successful and the values can...

Page 265: ...tion element S010 64 2 Establish 209 receptivity 3 Create an exposure 85kV 22 mAs 209 A message is displayed in the action window to indicate that the exposure is complete and the status indicator in...

Page 266: ...ment S030 11 The 5x5 volume is selected 2 Establish 209 receptivity 3 Create an exposure 85kV 0 34 mAs 209 A message is displayed in the action window to indicate whether adjustment has been successfu...

Page 267: ...status indicator turns green NOTE The calibration values are automatically saved 7 If volume 11x10 is available Click on the Next button or select the volume 11x10 in the structure tree S030 13 The 1...

Page 268: ...tree under 3D Geometry calibration 5x5 click on the Molar left element S010 25 S010 39 The Molar left element is selected 4 Establish 209 receptivity 5 Create an exposure 85kV 22 mAs 209 A message is...

Page 269: ...been successful and the status indicator in front of the 5x5 Premolar left element in the structure tree responds accordingly Calibration not OK status indicator stays red Calibration OK status indic...

Page 270: ...21 2 Attach the Mult O Meter sensor in the middle of the sensor unit flat panel detector side 3 Select a program from the 3D program group on the control panel The unit moves to the 3D exposure positi...

Page 271: ...the structure tree under Reset on the element Reset Pan Reset Ceph or Reset 3D depending on what data you would like to reset 2 To start resetting the adjustment and calibration data click on the Res...

Page 272: ...lorer or Firefox 2 In the address line enter https and the IP address of your unit Example https 192 168 15 240 Tip The IP address of your unit can be found on the info screen 3 Confirm your input wit...

Page 273: ...ting the touchscreen via the web service b A message is displayed in the web browser to indicate that the adjustment screen has been opened on the display The first adjustment screen appears on the co...

Page 274: ...ms GmbH 13 3 Adjusting the touchscreen via the web service Installation Manual ORTHOPHOS SL 8 Touch the center of the square in the bottom right corner of the screen 9 Touch the center of the square i...

Page 275: ...nt and calibration Installation Manual ORTHOPHOS SL 13 3 Adjusting the touchscreen via the web service b The unit switches to PC service mode Service mode is displayed on the Easypad via the PC servic...

Page 276: ...oard following replacement of the board This is possible via the Restore sensor data menu see the Service Manual Open the Backup Restore sensor data Open the backup restore data menu The unit and comp...

Page 277: ...4 1 Save sensor data b 5 Click on the Service button in the header line of the main menu The dialog box for entering the access data opens 6 Enter the following access data Username service Password s...

Page 278: ...or data element The Backup sensor data menu is displayed 2 Click on the Backup sensor data button The Download Backup menu opens A selection dialog box opens in your browser 3 In the selection dialog...

Page 279: ...Opening Unit information b 14 2 Opening Unit information SIDEXIS 4 is installed The workstation software for the corresponding X ray component or sensor is installed on the workstation 1 Start SIDEXI...

Page 280: ...016 14 Final work Sirona Dental Systems GmbH 14 2 Opening Unit information Installation Manual ORTHOPHOS SL The Device configuration menu opens 4 Click on the Device details button in the structure tr...

Page 281: ...ona Dental Systems GmbH 14 Final work Installation Manual ORTHOPHOS SL 14 2 Opening Unit information b The Unit information Advanced information is displayed 5 Click on the Downloading button to load...

Page 282: ...Operating system according to the op systems approved in the installation instructions Removable disk drive1 DVD ROM DVD RAM 1 supported by SIDEXIS function required only if not performed by an additi...

Page 283: ...Installation Manual ORTHOPHOS SL 14 4 Unit handover b 14 4 Unit handover Unit handover Give the customer all of the patient positioning aids test phantoms and special tools included in delivery inclu...

Page 284: ...t K p a p d e OMNS p W m d aPSPOKMPNKMNKMOKMO MPKOMNS KJk KW NON UNO p a p d e rp W c PN aJSQSOR _ d p a I f K QUPR p a I k OUOTP rp l k SQ VR NQO aPSPO K K...

Reviews: