Your Scientemp freezer has been pre-programmed at the factory and ready for use.
You only have to:
1) Plug your freezer into its own dedicated circuit and
2) Set your desired temperature on the controller
~~
To set the desired set point temperature, press and hold the * button, then while holding down the * press the
UP arrow to increase the set point temperature or press the DOWN arrow to decrease the set point temperature.
3) (If your freezer is equipped with a chart recorder) Replace chart paper on
recorder and remove black rubber cap on red recorder pen.
~~Press and hold the “change chart” button (#3) for approximately one second until pen begins to move to the left
of the chart and then release the button. Wait until the pen has moved completely off of the chart paper. To remove
the chart paper, unscrew (counter-clockwise) the knob at the center of the chart paper. Remove old paper and
position new one so that the correct time line coincides with the time line groove on the chart plate. Re-attach the
knob, screw securely (by hand). Press and hold change chart button (#3) again until pen begins to move back and
then release button.
•
Provide at least 3” of space around the freezer for ventilation. Avoid placing
the freezer in locations exposed to direct sunlight, heat registers or any other
heat source.
•
Alarm On/Off switch should remain in the “Off” position until the freezer
reaches desired set point temperature. When the temperature has been
reached, turn the alarm switch to the “On” position.
•
Cabinet should not be loaded with product until the cabinet has operated for
24 hours at the desired set point temperature.
Maintenance tips for your Scientemp Freezer:
•
Regularly remove any frost build up by using a plastic/wooden scraper.
•
Check and replace battery on Chart Recorder (if equipped) on a regular
bases.
•
The black finned condenser in the refrigeration compartment of the freezer,
should be cleaned of dust and debris about once a year (or more often if
freezer is located where excessive dust and dirt may occur)
~~This can be done by vacuuming carefully or by using forced air to blow the debris off.
3565 S. Adrian Hwy ~ Adrian, MI 49221-9253 ~ 1-517-263-6020 ~ fax 517-263-5492 ~ toll-free 800-968-2653
S C I E N T E M P C O R P .
Medical Freezers ~ Pharmacy Freezers ~ Culture Storage Freezers ~ Blood Storage/Blood Plasma Freezers
Adhesive Storage Freezers ~ Industrial/Manufacturing Freezers ~ Flammable and Explosion Proof Freezers