TH90K(E) Multi-function Infrared Thermometer
Functions
Room Temperature
Suitable ambient temperature is important for the baby and patient, the thermometer always help you recognize the
room temperature.
When in power off, room temperature will always be shown on the screen with
icon.
Please see the “Use of the thermometer” section to learn how to get the correct room temperature.
How to turn off Room Temperature: When in power on, press and hold the “ON/MEM” button for 5 sec. until
“
OFF”
will be shown on the screen to release the button.
Ear Temperature
The thermometer has been designed for practical use. It’s not meant to replace a visit to the doctor. Please also
remember to compare the measurement result to your regular body temperature.
Please see the “Use of the thermometer” section to learn how to measure the body temperature.
Forehead Temperature
The thermometer has been designed for practical use. It’s not meant to replace a visit to the doctor. Please also
remember to compare the measurement result to your regular body temperature.
Please see the “Use of the thermometer” section to learn how to measure the body temperature.
Surface Mode
The surface mode shows the actual, unadjusted surface temperatures, which is different from the body temperature.
It can help you to monitor if the object temperature is suitable for the baby or patient, for example the baby’s milk.
Please see the “Use of the thermometer” section to learn how to measure the object temperature.
Fever Alarm
If thermometer detects a body temperature
≧
37.5°C, there will be a long beep sound followed by three short beep
sound to warn the user for potential fever.
Last Reading
When you get a new temperature reading in ear mode, the last reading will be shown on the screen (at top right
corner) with the
icon.
Memory Locations
There are total 25 sets memories for ear measurements.
When in power on, Press the “ON/MEM” button to see the temperature stored with
icon.
°C / °F Switch
In “power off” mode (The LCD only shows the room temperature). Press and hold the “SCAN” button, then
immediately press and release the “ON/MEM” button. After 3 seconds, icon “°C” will be switched to icon “°F”. (You
may use the same way to switch back to “°C”)
Use of the thermometer
1. Always make sure the probe is clean, and without damage.
2. Power on: Press “ON/MEM” button.
3. Mode selection:
3.1
Ear mode
: Press the “ON/MEM” button, push up the raised brim on the
sliding panel on the above of LCD screen(see figure 4)
until
hearing a click sound
,the mode of thermometer is ear mode.
In this mode,
please press the “SCAN” button (see figure 5)
to get the body temperature by ear measurement.(see figure 1)
3.2 Forehead mode
: Press the “ON/MEM” button , push back the raised brim
on the sliding panel on the above of LCD screen(see figure 6),
the mode of thermometer is forehead mode.
In this mode, you can get the body temperature by forehead
measurement.(see figure 2)
Specifications
Temperature measurement range:
Ear/Forehead mode: 34~42.2°C (93.2~108°F), Surface mode: -22~80°C (-7.6~176°F)
Operating temperature range: 10~40°C (50~104°F)
Storage temperature Range: It should be stored at room temperature between -20~+50°C,
RH
≦
85%
Transportation temperature shall be less than 70°C, RH
≦
95%
Comply with ASTM E1965-98, EN12470-5:2003 Clinical thermometers-Part
5:Performance of infra-red ear thermometers(with maximum device),
IEC/EN60601-1-2(EMC), IEC/EN60601-1(Safety) standards.
Accuracy:
For ear mode: +/-0.2°C (0.4°F) within 35.5~42°C (95.9~107.6°F),
+/-0.3°C (0.5°F) for other range.
For forehead mode: +/-0.2°C (0.4°F) within 36~39°C (96.8~102°F),
+/-0.3°C (0.5°F) for other range.
For surface mode: +/-0.3°C (0.5°F) within 22~42.2°C (71.6~108°F),
/-4% or +/-2°C (4°F) whichever is greater.
Fever alarm & Memory under “ Ear/Forehead mode ”
°C / °F Switch mode
Battery: one lithium cell battery (CR2032 *1pcs)
This thermometer converts the ear/forehead temperature to display its “oral equivalent.”
:LED Backlight (optional)
LED Back
l
ight will automatically turned on after measurement, and automatically turned
off after 5 seconds
Do not use this device for Live & Dead decision or Safety related applications.
Please
consult with doctor if you have health concerns.
There is no gender and age limitation for using infrared thermometer.
This is not an AP
or APG product.
Figure 1 Figure 2
Figure 3
Figure 6
Figure 4
Ear Mode
Forehead Mode
Figure 5