MicroPurge® Flow Cell Quick Start Guide 
MP25 & MP25T 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Set up Flow Cell: 

1.

 

Remove storage cup from sonde 

2.

 

 For MP25, attached flow cell adapter to the threads on the sonde bottom cap. 

3.

 

Insert sonde/adaptor into flow cell and push into bayonets slots, then turn sonde clockwise to lock. 

4.

 

Determine necessary flow cell inlet connection fittings based on pump discharge tubing type and size. 

5.

 

Connect the included flow cell flexible tubing from pump to flow cell and then from flow cell discharge to purge 
water collection bucket. 

 

Connect to App: 

1.

 

If not already installed, download and install the MicroPurge Android app on your Android device (tablet, phone 
or Android OS laptop) 

2.

 

Enable the Bluetooth connection on your device for pairing. 

3.

 

Locate the Bluetooth ID on the bottom of the MP25 external battery pack 

4.

 

Launch the MicroPurge and watch for the Bluetooth ID to appear on the screen after 1-2 minutes. 

5.

 

Select the ID on the screen to connect the sonde to your device. 

6.

 

After the Sonde has successfully connected, press Settings, MicroPurge Sonde then at the top of the list press 
Calibrate. 

 

To Calibrate: 

See opposite side for 

Quick Start Calibration Guide

 

To Start PurgeScan Logging 

1.

 

Select the desired parameters to be measured from the Purge Scan > Sensors & Parameters menu  and hit Save 

2.

 

Select desired Stabilization parameters  from the Purge Scan > Stabilization Criteria menu and set the desired 
stabilization range for each parameter in percent or units 

3.

 

Select Logging Interval using the pulldown menu from 1-10 minutes 

4.

 

If desired, enable the Trending function using the on/off toggle button on the Stabilization Criteria screen 

5.

 

Save Stabilization parameters and Logging Interval. 

6.

 

Create PurgeScan log name or select from an existing file. 

7.

 

Hit OK 

8.

 

From the Tablet/Phone menu, look to see where the PurgeScan log files are being saved. “Using Phone Storage” 
will save files to the internal device memory; touch this to toggle to switch to “Using SD card if available” to save 
files to an SD card in the device. 

9.

 

If backup logging to the sonde is desired, go to the MicroPurge Sonde > Backup Log Setup and create Backup Log 
file name or select from an existing file. 

10.

 

Hit OK 

11.

 

 Turn PurgeScan Log ON and then, in the pop-up window, hit OK 

12.

 

If Stabilization function is desired, hit OK in the Stabilization pop-up window. 

13.

 

After stabilization has been achieved and purging is completed, go to Settings and select “Turn PurgeScan Log 
OFF” and then hit “OK” in the pop-up window.  

14.

 

To View your saved PurgeScan Log files, hit the View PurgeScan Logs button on the main screen. Select the 
desired log file name using the pull-down menu arrow in the upper right corner 

 

 

 

 

 

 

 

MicroPurge Application 

tinyurl.com/ycal8zyq

 

MP25 & MP25T Manual 

tinyurl.com/yawc4kb7

 

MP25 & MP25T Quick 

Calibration Guide 

tinyurl.com/y8lfh8j8

 