background image

User Manual

•  When there is suboptimal contact between the scalp and 

the disks, the Applicator will beep. The cause is usually 
insufficient pressure, having too few disks touch the scalp, 
or too much hair under the disks.

•  When there is good contact with the scalp, the Applicator will 

emit red laser light (LLLT) from between the disks. The red 
light will be turned off when contact is poor.

•  If the Applicator beeps a “poor contact” warning when moving 

the Applicator quickly over the scalp, try slowing down to a 
stop with the disks touching the scalp, until the beeping stops. 
Also try increasing the pressure of the device on the scalp.

•  Beginners are encouraged to use stimulation profile I, and 

after a couple of weeks proceed to stimulation profile II. If 
the intensity and duration are too demanding, stay a while 
longer on profile I.

•  Profile III is intended for users with long or dense hair. For 

such users, it is recommended to part the hair and roll back 
and forth across the parting line for half a minute, then move 
on to another parting line.

REPLACING  THE CARTRIDGE

The disks that come in contact with the scalp are for personal use. 
With time and usage, their performance degrades. The Cartridge 
usage is therefore monitored and the user is alerted when it 
needs to be replaced.

•  When the Cartridge exceeds 80% of its designed lifetime, a 

“Replace Cartridge” alert will flash on the Base.

REMOVING THE CARTRIDGE

•  Hold the Applicator with one hand and the grooves at the 

Cartridge’s end with the other.

•  Pull out firmly the Cartridge.
•  Push and click a new Cartridge into place.

MAINTENANCE

•  Keep all parts of the device clean and dry.
•  The Base and Applicator should be cleaned with a damp cloth 

and then dried.

•  The disks and the area between them can be cleaned with a 

brush. For sanitary purposes, roll the disks on an alcohol pad, 
and dry before use. 

DISPOSAL AND RECYCLING

•  The Applicator contains a rechargeable battery - please dispose 

of it only in accordance with local regulations.

•  To remove the rechargeable battery, use a flat screwdriver to 

pry open the rear panel of the Applicator.

•  Please follow all relevant regulations when disposing of the 

other parts of the device. 

WARNINGS

•  Keep the device away from children.
•  Use the device only according to instructions.
•  Use only on scalp - do not use on other body parts.
•  Do not use on wet or damp hair. The hair and scalp should be dry.
•  Do not use the Applicator without a properly attache Cartridge. 

Do not remove the Cartridge while the Applicator is turned on.

•  Do not shine the red laser light into eyes.
•  Keep the device away from water or wet areas.
•  Store and operate the device in normal room temperature.
•  Avoid use if device was dropped, physically damaged or wet.
•  In case of malfunction, contact the seller - do not try to fix the 

problem yourself.

SPECIFICATIONS

Input

•  4.5 to 6 V DC
•  Power consumption: up to 60mA when not charging, up to 

500mA when charging.

•  Interface: USB2

Output

•  Current: up to 300mA, 0~100hz
•  Voltage: up to 30V
•  Laser: up to 5mW, 650nm

Distributions: