background image

USER GUIDE

Blood Glucose Monitoring System

Activa

Summary of Contents for Smartest Activa

Page 1: ...USER GUIDE Blood Glucose Monitoring System Activa...

Page 2: ...lution Test Blood Testing 11 Running a Blood Glucose Test 12 Recalling the Memory and Viewing the Average 13 Expected Values Reference 14 Transmitting Results Maintenance 15 Maintenance of Your System...

Page 3: ...rm in this system can only be used during steady state blood glucose conditions contains a speaking function that provides audible messages to assist the user however the system is not intended for vi...

Page 4: ...rile Lancets 8 Lancing Device 9 Package Inserts for Test Strips Control Solution and Lancets Lancing Device A single level control solution Medium is provided in the Starter Kit and no control solutio...

Page 5: ...st using symbols and simple messages 3 M Key Power ON OFF also for memory recalling mode please refer to manual for detailed function description 4 C Key Confirm setting view average of memory and fla...

Page 6: ...ingertip When to use AST Medication stress illness food and exercise can affect blood glucose levels Capillary whole blood at the fingertip can reflect test changes faster than capillary blood at othe...

Page 7: ...you to get a drop of blood for AST 2 Insert a new lancet firmly into the lancet holder 3 Twist off the protective top of the lancet 4 Close the end cap of the lancing device and grasp the lancing dev...

Page 8: ...ip out of the vial and quickly cover the vial to keep air out 8 Insert a test strip into the meter while the meter is off The meter will turn on with a blinking blood drop icon appears on the LCD scre...

Page 9: ...site a little longer before using the lancing device 11 Apply your blood to the front edge the tip of the test strip and the meter will start counting down for 6 seconds then display the test results...

Page 10: ...st Activa meter that has the same reference number Please keep the test strip vial away from children The test strips and vial cap can present a choking hazard Please be cautious when removing the lan...

Page 11: ...and let a blood drop form Please see Figure 6 4 5 If you wish to flag the glucose test as a Before Meal test after you hear Apply blood now press the C button two 2 times You will then hear Ready for...

Page 12: ...n the right side of the reading as an After Meal test and a b as a Before Meal test Please see Figure 6 7 8 Remove the test strip from the meter and the meter will turn off automatically You will hear...

Page 13: ...a beep sound if the battery is installed correctly Please see Figure 7 1 Please see Figure 7 2 Please see Figure 7 3 Important 1 It is suggested that you replace the batteries when battery icon appea...

Page 14: ...sappear from the LCD screen If you press Up button from 0 the voice volume will be adjusted upward by one level each time you press the Up button You will hear a beep sound to indicate the current voi...

Page 15: ...talking meter is voice ON If you do not press any thing for 2 seconds the meter will say Goodbye and the LCD screen will display OFF Below the Up and Down buttons there is a round button which is for...

Page 16: ...1 Day 12 Hour 13 Minute 1 Measurement Units 6 Battery Symbol 3 Setting 2 Memory 10 Month 11 Day 12 Hour 13 Minute 4 A b C 5 Blood drop icon 7 Voice Symbol 8 Day Average 9 Alarm Reminder By pressing th...

Page 17: ...hear The year is xxx if the voice is on Press C key to confirm and then the month segment on the LCD will flash You will hear Please set the month Please see Figures 9 2 9 3 9 4 9 5 3 Set the Month Pr...

Page 18: ...e Hour Press the Up or Down key on the right side until the desired hour appears on the LCD screen Each time you adjust the hour you will hear The hour is xx am or pm if the voice function is on Press...

Page 19: ...9 11 Figure 9 12 8 Set the measurement unit Press Up or Down to toggle between the selection of mg dL and mmol L The LCD screen will display either mg dL or mmol L when pressing Up or Down key Each t...

Page 20: ...manage your diabetes If you live in U S A you should use mg dL your results will never have a decimal point If you live in Canada or some European countries you should use mmol L your result will alw...

Page 21: ...e 10 1 C 3 Squeeze a small amount of control solution on a flat surface and apply to the front edge of the test strip then you will hear Testing in process 4 Wait for 6 seconds and the meter will disp...

Page 22: ...nt measurement ranges 5 Shake the vial discard the first drop of control solution and wipe off the dispenser tip to ensure a good sample and an accurate result 6 Use only for 3 months after first open...

Page 23: ...tective tip of the lancet 4 Close the end cap of the lancing device and grasp the lancing device as shown Slide into locking position 5 Wash your hands in warm soapy water Be sure to rinse and dry wel...

Page 24: ...use disposable lancet ready for taking the blood sample 8 You can start using the lancing device or the disposable lancet to take the blood sample from the fingertip When a blinking blood drop icon ap...

Page 25: ...for 6 seconds and then display the test results on the screen You will hear The glucose level is number unit For those flagged glucose tests the screen will display an A on the right side of the read...

Page 26: ...2 2 1 With meter off press M key to turn on the meter after a brief full screen display and you will hear Recalling the memory the meter will display the total number of test data stored in memory The...

Page 27: ...out exit the memory mode If you want to exit the memory operations the meter can be turned off by pressing and holding the M key for 2 seconds or the meter will automatically turn off after 20 seconds...

Page 28: ...hear Goodbye and the LCD screen will display OFF 3 If you want to view the average results for Before Meals and After Meals for a specific day period 7 14 28 60 90 days please see below A Press the Up...

Page 29: ...ations The 7 14 28 60 90 days average is the average of test results from the last 7 14 28 60 90 days respectively Note 1 The out of range HI LO test results will not be stored in the memory 2 Do not...

Page 30: ...personal computer Please follow the software instructions for data transmission 4 When all results have been transmitted or when you press and hold the M key for 2 seconds to exit the transmission mo...

Page 31: ...clean surface of the meter with a soft cloth that has been slightly dampened with alcohol 3 4 If this meter is used by a healthcare professional infection control policies should be strictly followed...

Page 32: ...e is above its operating range Audible message The temperature is too high E_3 Battery Low No further tests can be performed Audible message Battery low E_4 The blood sample is not sufficient Audible...

Page 33: ...eter is operating in an ambient temperature above 40 or 104 which is the cap of operating temperature System operation range is 10 40 50 104 Repeat the test when the temperature is within the range E_...

Page 34: ...tery power or replace them if necessary 4 Please call Customer Service The meter does not count down The meter is defective or the battery is low Replace the battery If the meter fails to operate call...

Page 35: ...dical issues or advice For questions related to your health condition please call your doctor or healthcare professional Smartest Activa Warning Your meter can show results either in mg dL or mmol L u...

Page 36: ...lood may cause incorrect results Uric acid 10 9 mg dL acetaminophen 6 2 mg dL ascorbic acid 4 5 mg dL ibuprofen 37 5 mg dL 6 The test strips may be used at altitudes up to 5 280 feet 1 609m without an...

Page 37: ...600 mg dL 1 1 33 3 mmol L Measuring Time 6 seconds Code required No code device Batteries ALKALINE 24A LR03 AAA 1 5V X 2 Operating Range 10 40 50 104 Humidity 20 80 RH Width 54 3 mm Length 100 4 mm T...

Page 38: ...so find the information on how the system works and on the performance characteristics 20 Ordering Supplies To order supplies please contact Progressive Health Supply Source Corp One Industrial Way We...

Page 39: ...Building E Eatontown NJ 07724 USA TEL 1 888 88PHSSC Hours 9 AM to 5 PM EST Mon to Fri Outside of these times contact your healthcare professional Reference 1 2 American Diabetes Association Diabetes...

Page 40: ...Progressive Health Supply Source Corp One Industrial Way West Building E Eatontown NJ 07724 USA www progressivehealthsupply com TEL 1 888 88PHSSC...

Reviews: