Pump Removal and Replacement  

Procedure for the  Instrument 115/230VAC

 

 
Pump Removal 
 
TURN THE MAIN POWER SWITCH LOCATED AT THE LEFT SIDE OF THE LOWER CABINET 
TO THE OFF POSITION. 
 

 

 
Turn the pump switch on the head to the off position (to the left ¼ turn).  Next you must trip breaker 
(down position) inside the lower cabinet electrical assembly as shown below. 
 

  

  

 

 
Remove 4 screws from pump electrical assembly and remove assembly cover.  Retain these 4 screws for 
reassembly. 
 

  

   

Summary of Contents for Probrix+

Page 1: ...LOCATED AT THE LEFT SIDE OF THE LOWER CABINET TO THE OFF POSITION Turn the pump switch on the head to the off position to the left turn Next you must trip breaker down position inside the lower cabin...

Page 2: ...nnections Take a 7 mm driver and remove nuts securing both the hot and neutral wires Remove wires from terminals and replace nuts and tighten so nuts are not lost Unscrew ground wire and replace screw...

Page 3: ...ctrical assembly along with the 4 screws and tighten Remove pump from lower body Now you must remove the pump from the mounting bracket Remove the 4 nuts bolts and washers from the rear of the mountin...

Page 4: ...rs and nuts retained from the removal procedure Remove 4 screws from replacement pump electrical assembly and remove assembly cover Retain these 4 screws for reassembly Place pump in lower assembly of...

Page 5: ...r the pump Feed the power cord into the pump electrical assembly as shown below and tighten cable gland Remove nuts from the neutral and hot terminals Place the eyelets for each lead onto the proper t...

Page 6: ...he on position up position inside the lower cabinet electrical assembly Close lower cabinet electrical assembly and secure Turn the pump switch on the head to the on position to the right turn until i...

Page 7: ...on the capacitor assembly as there are wires running from the pump to this assembly Terminal connections are shown below NEW STYLE Jumpers are on V1 U2 upper W1 V2 lower Motor rotating CCW For CW rota...

Page 8: ...Version 01 23 2003 Page 8 of 8 AC power will be connected to terminals W1 left V2 right W 1 V 2...