VISERA UHD 4K Camera System Quick Reference Guide

Disclaimer:

 Information provided in this guide is not exhaustive. Please refer to the complete Instructions For Use for each 

item of equipment which should be thoroughly read and understood prior to handling, use and storage of the equipment. 

Light Source

Camera Head

Switch On

Switch on the camera system using the workstation master 
switch which will then turn from white to green.

There is a secondary power switch on the transformer 
under the workstation. This should remain switched on.

Insert the camera head into the camera processor, 
ensuring the connector is the correct way up.

The colour bars on the monitor will now be replaced with 
an image from the camera head.

Optical connecting part

UP mark

Video connector

Electrical contacts

Video connector socket

Video connector

UP mark

If required, place the lamp onto standby by pressing and 
holding the lamp button.

Press the brightness mode button to select Auto.

Warning: 

When turning ON the light source, never allow 

the distal end of an endoscope or a light guide cable to 
come in contact with the patient and/or other flammable 
materials, such as operating room drapes. Patient injury 
and/or a fire can result.

Connect the light guide cable to the output.

Output socket

Brightness mode 
indicators

Brightness mode 
button

Lamp indicators

Lamp button

Emergency 
lamp

Control panel

 

Issue 1 

05/02/2021

 

KMF-08930 

NCDR 02868

Summary of Contents for VISERA

Page 1: ...or is the correct way up The colour bars on the monitor will now be replaced with an image from the camera head Optical connecting part UP mark Video connector Electrical contacts Video connector socket Video connector UP mark If required place the lamp onto standby by pressing and holding the lamp button Press the brightness mode button to select Auto Warning When turning ON the light source neve...

Page 2: ...ly enclosed Press and hold the white balance button on the processor until White Balance Complete appears on the monitor The white balance status changes to completed White balance window White balance window Warning Do not use the high intensity mode when connecting to an endoscope that has an insertion tube outer diameter of 4 1 mm or less The endoscope becomes hot while the lamp is lit up and c...

Page 3: ...ing the monitor which can be switched off by accident Ensure monitor input is set to SDI 1 by using the monitor input buttons Unplug camera head colour bars should appear If no colours appear contact Olympus Service Helpdesk Dark image Check condition of telescope Check condition of light lead High Intensity Mode Doesn t light up Check a compatible Olympus light lead is being used Main examination...

Page 4: ...ical Europe OlympusMedUKIE KeyMed House Stock Road Southend on Sea Essex SS2 5QH UK Olympus Medical UKIE Olympus Medical Europe OlympusMedUKIE For further information please contact Olympus Customer Services on 01702 616 333 or visit www olympus co uk Issue 1 05 02 2021 KMF 08930 NCDR 02868 ...

Reviews: