NOVUS AUTOMATION 

1/7 

 

 

Indicator N1500LC 

UNIVERSAL INDICATOR – INSTRUCTIONS MANUAL – V2.3x E 

SAFETY ALERTS 

The symbols below are used on the equipment and throughout this 

document to draw the user’s attention to important operational and 

safety information. 

 

 

CAUTION: 

Read this manual carefully prior to 

installation and operation of the 

unit. 

CAUTION OR DANGER: 

Electric shock hazard 

All safety related instructions that appear in the manual must be 

observed to ensure personal safety and to prevent damage to either the 

instrument or the system.  If the device is used in a manner not 

specified in this manual, its safety features may be impaired. 

INTRODUCTION 

The N1500LC is an indicator designed for weighing system with load 

cells. It accepts a wide variety of electrical signals. It has a display 

with six-digit LED display for the measured value and device 

parameters. 
The whole device configuration is done through the keyboard without 

any change in the circuit. Thus, the selection of input type and the 

type of alarms action, besides other special functions are all 

accessed and programmed through the front keypad. 
It is important that you read this manual carefully before using the 

instrument. This is an electronic device that requires care in handling 

and operation, and if properly used, it will be very efficient in the work 

required. 
The main features of the basic version of N1500LC are as follows: 

 

Input: 4-20 mA, 0-20 mA, 0-50 mV, 0-20 mV e -20 a 20 mV; 

 

Increasing or decreasing display; 

 

10 Vdc (or 5 Vdc) power supply for load cells; 

 

Memory for maximum and minimum values; 

 

Hold, peak hold, tare, zero tare and automatic zero functions;  

 

Digital input; 

Optionally can have: 

 

Process Variable (PV) retransmission in 0-20 mA ou 4-20 mA; 

 

RS485 MODBUS RTU serial communication; 

 

3rd and 4th alarm relays. 

 
 
 
 

The front panel is shown below, with a description of its elements. 

 

Figure 1 - 

Identification of the elements of the front panel 

Display

: Displays the value of the measured variable (PV) and the 

mnemonics of the device programming parameters. 

A1, A2, A3 

and

 A4

: show active alarms. 

Rx

 and 

Tx

: indicate RS485 communication line is active. 

 

P

 

key 

- This key is used to access different displays with the 

programmable parameters of the device. 

 

BACK  key

  -  This key is used to go back to the previous 

parameter displayed in the menu display. 

 

INCREMENTS/ZERO key and 

 

DECREMENTS/ZERO 

key 

-

 

They make possible the change the parameter values. 

They are also used to display maximum and minimum 

values stored in memory. 

 

Key Special FUNCTION 

- This special function key is used for 

programmed functions as explained in the SPECIAL FUNCTION KEY 

section of this manual. 

PROCESS VARIABLE INPUT - PV 

The type of input to be used by the indicator must be set by the user 

via the keyboard, among types established in 

Table 1

All input types available are factory calibrated and require no further 

adjustment by the user, except the definition of range of indication. 

Type 

Code 

Measurement Range 

Non-linear 4-20 mA 

c.4-20 

Programmable indication range. 

Three 

maximum

 range options: 

-32000 to +32000 

0 to +60000 

0 to +120000 (only even values) 

 

Nonlinear signals will be 

linearized according to the 

programmed linearization. 

Non-linear 0-20 mA  

c.0-20 

Linear 4-20 mA  

4-20 a 

Linear 0-20 mA 

0-20 a 

Non-linear 0-50 mV  

c. 50 

Non-linear -20 a 20 mV 

c. -20 

Non-linear  0-20 mV 

C. 20 

Linear 0 – 50 mV  

50 

Linear -20 a 20 mV  

-20 . 20 

Linear 0-20 mV  

20 

Table 1

 - Input types accepted by the indicator 

 
 
 
 

Summary of Contents for N1500LC

Page 1: ...utomatic zero functions Digital input Optionally can have Process Variable PV retransmission in 0 20 mA ou 4 20 mA RS485 MODBUS RTU serial communication 3rd and 4th alarm relays The front panel is sho...

Page 2: ...with value 0 zero The Warning LEDs associated with the alarm always light up when the alarm condition occurs regardless of the current state of the output relay which may be de energized momentarily...

Page 3: ...to adopt the Custom Linearization option This feature combines the input signal to 30 user defined line segments setting two points for each segment a start and end and the respective display values T...

Page 4: ...HE INDICATOR WORK LEVEL This is the first level At power up the indicator will display the Process Variable PV The alarm triggering points are also displayed at this level alarm Setpoints To advance i...

Page 5: ...et also obey these limits In lol Input Low Limit Determines the minimum limit for input signals The band created can be ascending or descending characteristic in relation to the input signal behavior...

Page 6: ...erly configured the instrument may show false error messages particularly those related to the type of input selected SPECIAL RECOMMENDATIONS Should the indicator be repaired some special handling car...

Page 7: ...e the serial communication interface Communication Baud Rate bavd parameter and Communication Address adres parameter SPECIFICATIONS DIMENSIONS 48 x 96 x 92 mm 1 8 DIN Approximate weight 250 g PANEL C...

Reviews: